IM 12D7B3-E-H
Unclassified Location
+
_
G
Classified Location
Protective
earth
Intrinsically safe design
FM Class I. Div. 2. Group ABCD: T4 for ambient temperature < 55 ºC
T6 for ambient temperature < 40 ºC
EXA SC202S analyser
see text below
For electrical data:
SENSOR(S)
terminal 11-16
Max. cable length: 60 mtr.
Cable diam.: 3 to 12mm
Classified Location
Unclassified Location
+
_
G
Protective
earth
+
_
Load
Resistance
FM approved power supply
V
OC
≤ 31.5 VDC
Intrinsically safe design
FM Class I. Div. 2, Group ABCD: T4 for ambient temperature < 55 ºC
T6 for ambient temperature < 40 ºC
EXA SC202S analyser
For electrical data:
see text below
SENSOR(S)
terminal 11-16
Max. cable length: 60 mtr.
Cable diam.: 3 to 12mm
FM approved power supply
V
OC
≤ 31.5 VDC
+
_
Figure 2
Figure 1
•
Electrical data of the EXA SC202S:
- Supply circuit (ter and -):
- Sensor input circuit (terminals 11 through 16):
Maximum input voltage Vmax = 31.5 V.
Maximum output voltage V
t
= 14.4 V.
Maximum input power P
i
= 1.2 W
Maximum output current I
t
= 12.8 mA.
Effective internal capacitance C
i
= 22 nF
Maximum allowed external capacitance C
a
= 1.4 µF.
Effective internal inductance L
i
= 22 µH
Maximum allowed external inductance L
a
= 900 mH.
•
The Hand Held Terminal must be FM Approved in case it is used in the classified location.
When installing this equipment, follow the manufacturers installation drawing. Installation shall be in accordance with Article
501.4(B) of the National Electrical Code (ANSI/NFPA 79).
Nonincendive field wiring may be installed in accordance with Article 501.4(B)(3)
•
Grounding shall be in accordance with Article 250 of the National Electrical code
WARNING
- Substitution of components may impair suitability for Division 2
- Do not remove or replace while circuit is live unless area is know to be non-hazardous
- Explosion Hazard – Do not disconnect equipment unless area is know to be non-hazardous
- Do not reset circuit breaker unless power has been removed from the equipment or the area is know to be non-hazardous
FM non-incendive diagrams for SC202S-N