3-16
IM MV1000-01E
• Number of set ponts > Mes val, True val
Press the
Input
soft key and enter the value.
For
Mes val
(measured value), set a value that is greater than the previous value.
Press the
Measure
soft key to set the current measured value to
Mes val
. If you press
the
Measure
soft key when you are setting the calibration function for more than one
channel, the measured value of the channel set to
Frst-CH
is used as the
Mes val
for
all channels.
Range of Selectable Values for Mes val and True val
• On Channels Set to Lnear Scalng
–30000 to 30000 (decimal place is the same as that set for the scale)
• On All Other Channels
The measurable range
Example: If the range is 2 V, –2.0000 to 2.0000
Note
• If you change the Mode or Range settings, calibration correction (the
Number of set ponts
setting) is switched Off.
• You cannot set calibration correction for a channel that is set to Skip.
3.9 Calbratng Input Values (/CC1 Opton)