background image

4/12

IM MT300-71EN

6. 

When the digital multimeter display stabilizes, use the 

arrow

 (▲, ▼, ◄, and ►) keys to input the value 

displayed on the digital multimeter.

   

7. 

Press 

ENTER

. The full-scale positive adjustment value is set.

8. 

Press the 

MinusFull

 soft key.

   

9. 

On the standard DC voltage current source, output −5 V.

10. 

When the digital multimeter display stabilizes, use the 

arrow

 (▲, ▼, ◄, and ►) keys to input the value 

displayed on the digital multimeter.

   

11. 

Press 

ENTER

. The full-scale negative adjustment value is set.

Adjusting the Current Measurement

12. 

Press the 

Range

 soft key to select the +20 mA range.

   

13. 

Connect this instrument, digital multimeter, standard DC voltage current source, and standard resistor.

For the wiring method and instrument setup, see “Explanation.”

14. 

Press the 

PlusFull

 soft key.

   

15. 

On the standard DC voltage current source, 20 mA.

16. 

When the value displayed on the digital multimeter stabilizes, divide the displayed value by the standard 
resistor value to calculate the current.

17. 

Use the 

arrow

 (▲, ▼, ◄, and ►) keys to enter the calculated current.

   

18. 

Press 

ENTER

. The full-scale positive adjustment value is set.

19. 

Press the 

MinusFull

 soft key.

   

20. 

On the standard DC voltage current source, output −20 mA.

21. 

When the value displayed on the digital multimeter stabilizes, divide the displayed value by the standard 
resistor value to calculate the current.

Summary of Contents for MT300

Page 1: ...The following options are installed in the MT300 D A output adjustment DA option Voltage current measurement adjustment DM option Equipment Used Digital multimeter 3458A Keysight or equivalent Standa...

Page 2: ...ess the Range soft key to select the 2 V range Output value 4 Press the Zero soft key The zero point 0 V is output and the input screen appears 5 When the digital multimeter display stabilizes use the...

Page 3: ...soft key The voltage current measurement adjustment menu appears Note If the warning code 59 The measured value is out of range appears when you set an adjustment value the value may be exceeding the...

Page 4: ...A range 13 Connect this instrument digital multimeter standard DC voltage current source and standard resistor For the wiring method and instrument setup see Explanation 14 Press the PlusFull soft key...

Page 5: ...and 20 mA ranges are saved and the screen returns with adjustment menu Ending Adjustment Mode 1 On the adjustment menu press ESC 2 Press the OK soft key The screen returns the top menu Canceling the A...

Page 6: ...n the Next soft key The following menu appears 3 Press the Maintenance soft key The Maintenance menu appears 4 Press the Calibration Date soft key The following menu appears 5 If you calibrated the D...

Page 7: ...digital multimeter s voltage input terminal and the GS200 standard DC voltage current source s output terminal SOURCE GS200 H L MT300 V DMM 3458A H L A COM Set the 3458A measurement range to 10 VDC ra...

Page 8: ...Sets the full scale negative adjustment value of the voltage current measurement 10 CALibrate DMM ADJust PFULl Sets the full scale positive adjustment value of the voltage current measurement 10 CALi...

Page 9: ...te DA POINt command to set the adjustment point to PFULL If the adjustment point is not PFULL a 221 Setting conflict error will occur CALibrate DA ADJust ZERO Function Sets the zero point adjustment v...

Page 10: ...t target to DA If the adjustment target is not DA a 221 Setting conflict error will occur When executed the adjustment target is set to NONE CALibrate DMM ADJust MFULl Function Sets the full scale neg...

Page 11: ...ALibrate DMM POINt PFULL MFULL CALibrate DMM POINt PFULL Full scale positive value MFULL Full scale negative value Example CALIBRATE DMM POINT PFULL CALIBRATE DMM POINT CALIBRATE DMM POINT PFULL Descr...

Page 12: ...sure to write the adjustment values using the appropriate command before changing the adjustment target On models without the DM option sending a CALibrate FUNCtion DMM command will cause a 141 Inval...

Reviews: