11/12

IM MT300-71EN

:CALibrate:DMM:DATE?

Function 

Queries the date when the adjustment value of 
voltage/current measurement was written.

Syntax

 

:CALibrate:DMM:DATE?

Example

  :CALIBRATE:DMM:DATE?

 

-> :CALIBRATE:DMM:DATE "2019/09/12"

Description  On models without the /DM option, a “241: 

Hardware missing” error will occur.

:CALibrate:DMM:INITialize

Function 

Initializes the adjustment value of voltage/current 
measurement.

Syntax

 

:CALibrate:DMM:INITialize

Example

  :CALIBRATE:DMM:INITIALIZE

Description  •  Initializes the adjustment value of voltage/

current measurement. On models without the /

DM option, a “241: Hardware missing” error will 

occur.

 

•  Use the 

:CALibrate:FUNCtion

 command in 

advance to set the adjustment target to DMM. 

If the adjustment target is not DMM, a “221: 
Setting conflict” error will occur.

:CALibrate:DMM:MEASure?

Function 

Queries the measured voltage/current.

Syntax

 

:CALibrate:DMM:MEASure?

Example

  :CALIBRATE:DMM:MEASURE?

 

-> :CALIBRATE:DMM:MEASURE 5.0000E+00

Description  •  On models without the /DM option, a “241: 

Hardware missing” error will occur.

 

•  If the value is +OverRange or +OverFlow, 

9.90E+37 is returned. 

 

If the value is −OverRange or −OverFlow, 
−9.90E+37 is returned. 

 

If DMM is not selected with the 

:CALibrate:FUNCtion

 command, 9.91E+37 

is returned.

:CALibrate:DMM:POINt

Function 

Sets or queries the adjustment point of the 
measurement range of the voltage/current 
measurement to be adjusted.

Syntax

 

:CALibrate:DMM:POINt {PFULL|MFULL}

 :CALibrate:DMM:POINt?

 

PFULL: Full-scale positive value

 

MFULL: Full-scale negative value

Example

  :CALIBRATE:DMM:POINT PFULL

 :CALIBRATE:DMM:POINT?
 

-> :CALIBRATE:DMM:POINT PFULL

Description  •  On models without the /DM option, a “241: 

Hardware missing” error will occur.

 

•  Use the 

:CALibrate:FUNCtion

 command in 

advance to set the adjustment target to DMM. 

If the adjustment target is not DMM, a “221: 
Setting conflict” error will occur. A query returns 

the current value.

:CALibrate:DMM:RANGe

Function 

Sets or queries the measurement range of the 
voltage/current measurement to be adjusted.

Syntax

 

:CALibrate:DMM:RANGe {VOLTage, 
<

Voltage

>|CURRent,<

Current

>}

 :CALibrate:DMM:RANGe?

 

<Voltage> = 5 (V)

 

<Current> = 20 (mA)

Example

  :CALIBRATE:DMM:RANGE VOLTAGE,5V

 :CALIBRATE:DMM:RANGE?

 

•  5 V range

 

-> :CALIBRATE:DMM:RANGE VOLTAGE, 
5.000E+00

 

•  20 mA range

 

-> :CALIBRATE:DMM:RANGE CURRENT, 
20.0E−03

Description  •  On models without the /DM option, a “241: 

Hardware missing” error will occur.

 

•  Use the 

:CALibrate:FUNCtion

 command in 

advance to set the adjustment target to DMM. 

If the adjustment target is not DMM, a “221: 
Setting conflict” error will occur. A query returns 

the current value.

:CALibrate:DMM:WRITe

Function 

Writes the adjustment value of voltage/current 
measurement.

Syntax

 

:CALibrate:DMM:WRITe

Example

  :CALIBRATE:DMM:WRITE

Description  •  On models without the /DM option, a “241: 

Hardware missing” error will occur.

 

•  Use the 

:CALibrate:FUNCtion

 command in 

advance to set the adjustment target to DMM. 

If the adjustment target is not DMM, a “221: 
Setting conflict” error will occur.

 

•  When executed, the adjustment target is set to 

NONE.

Summary of Contents for MT300

Page 1: ...The following options are installed in the MT300 D A output adjustment DA option Voltage current measurement adjustment DM option Equipment Used Digital multimeter 3458A Keysight or equivalent Standa...

Page 2: ...ess the Range soft key to select the 2 V range Output value 4 Press the Zero soft key The zero point 0 V is output and the input screen appears 5 When the digital multimeter display stabilizes use the...

Page 3: ...soft key The voltage current measurement adjustment menu appears Note If the warning code 59 The measured value is out of range appears when you set an adjustment value the value may be exceeding the...

Page 4: ...A range 13 Connect this instrument digital multimeter standard DC voltage current source and standard resistor For the wiring method and instrument setup see Explanation 14 Press the PlusFull soft key...

Page 5: ...and 20 mA ranges are saved and the screen returns with adjustment menu Ending Adjustment Mode 1 On the adjustment menu press ESC 2 Press the OK soft key The screen returns the top menu Canceling the A...

Page 6: ...n the Next soft key The following menu appears 3 Press the Maintenance soft key The Maintenance menu appears 4 Press the Calibration Date soft key The following menu appears 5 If you calibrated the D...

Page 7: ...digital multimeter s voltage input terminal and the GS200 standard DC voltage current source s output terminal SOURCE GS200 H L MT300 V DMM 3458A H L A COM Set the 3458A measurement range to 10 VDC ra...

Page 8: ...Sets the full scale negative adjustment value of the voltage current measurement 10 CALibrate DMM ADJust PFULl Sets the full scale positive adjustment value of the voltage current measurement 10 CALi...

Page 9: ...te DA POINt command to set the adjustment point to PFULL If the adjustment point is not PFULL a 221 Setting conflict error will occur CALibrate DA ADJust ZERO Function Sets the zero point adjustment v...

Page 10: ...t target to DA If the adjustment target is not DA a 221 Setting conflict error will occur When executed the adjustment target is set to NONE CALibrate DMM ADJust MFULl Function Sets the full scale neg...

Page 11: ...ALibrate DMM POINt PFULL MFULL CALibrate DMM POINt PFULL Full scale positive value MFULL Full scale negative value Example CALIBRATE DMM POINT PFULL CALIBRATE DMM POINT CALIBRATE DMM POINT PFULL Descr...

Page 12: ...sure to write the adjustment values using the appropriate command before changing the adjustment target On models without the DM option sending a CALibrate FUNCtion DMM command will cause a 141 Inval...

Reviews: