2/12
IM MT300-71EN
6.
Press
ENTER
. The adjustment menu appears.
The features that can be adjusted on this instrument appear.
Adjust the D/A output. Adjust the voltage/current
measurement.
Adjusting the D/A Output
1.
Connect this instrument to a digital multimeter.
For the wiring method and instrument setup, see “Explanation.”
2.
Press the
D/A
soft key. Either 2 V or 5 V is output, and the D/A output adjustment menu appears.
3.
Press the
Range
soft key to select the 2 V range.
Output value
4.
Press the
Zero
soft key. The zero point (0 V) is output, and the input screen appears.
5.
When the digital multimeter display stabilizes, use the
arrow
(▲, ▼, ◄, and ►) keys to input the value
displayed on the digital multimeter.
6.
Press
ENTER
. The zero point adjustment value is set.
7.
Press the
PlusFull
soft key. The full-scale positive value (+2 V) is output, and the input screen appears.
8.
When the digital multimeter display stabilizes, use the
arrow
(▲, ▼, ◄, and ►) keys to input the value
displayed on the digital multimeter.
9.
Press
ENTER
. The full-scale positive adjustment value is set.