:CALibrate:FUNCtion
Function
Sets or queries the adjustment target (voltage/
current measurement, D/A output).
Syntax
:CALibrate:FUNCtion {DMM|DA|NONE}
:CALibrate:FUNCtion?
DMM: Voltage/current measurement
DA: D/A output
NONE: Not selected
Example
:CALIBRATE:FUNCTION DMM
:CALIBRATE:FUNCTION?
-> :CALIBRATE:FUNCTION DMM
Description • If you change the adjustment target before
executing the
:CALibrate:DMM:WRITe
or
:CALibrate:DA:WRITe
command, the
adjustment values that you have set will
return to their original values. Be sure to write
the adjustment values using the appropriate
command before changing the adjustment
target.
• On models without the /DM option, sending a
:CALibrate:FUNCtion DMM
command will
cause a “141: Invalid character data” error.
• On models without the /DA option, sending a
:CALibrate:FUNCtion DA
command will
cause a “141: Invalid character data” error.
:CALibrate:PASSword
Function
Enters the password for switching to adjustment
mode.
Syntax
:CALibrate:PASSword {<
String
>}
<String>: Password
Example
:CALIBRATE:PASSWORD "1111"
Description You must enter a password to use the commands
in the CALibrate group.
Enter a character string for the password.
To exit from adjustment mode, enter an empty
string (two consecutive double quotation marks).