5-34
IM XL100C-E
Query
XO?
Example
Set the communication method to Ethernet.
XO ETHERNET
When switching to serial communications (RS-
232, RS-485, or USB)
Syntax
XO p1,p2<terminator>
p1
Serial communications
SERIAL
p2
Serial communication type
RS232
,
RS485
, or
USB
Query
XO?
Example
Set the communication method to RS-232.
XO SERIAL,RS232
When switching to serial communication printer
connection (RS-232PRINTER communication)
Syntax
XO p1,p2,p3 <terminator>
p1
Serial communication
SERIAL
p2
Serial communication type
RS232PRINTER
p3
Manual printout
OFF:
0
Measured data:
1
Screen data:
2
Measured/Screen data:
3
Query
XO?
Example
Set the XL100 to print only the measured data
during the manual printout.
XO SERIAL,RS232PRINTER,1
XZ
Selects the trigger
Syntax
XZ p1,p2<terminator>
p1
Trigger mode
Single:
0
Continuous:
1
p2
Pre-trigger and delay trigger setting
Pre-trigger:
1
Delay trigger:
2
p3
Pre-trigger or delay trigger count
0
to
600
Query
XZ?
Example
Set the trigger to single and the pre-trigger count
to 10.
XZ 0,1,10
XK
Selects the start trigger
When setting the start trigger to none
Syntax
XK p1
<terminator>
p1
Start trigger type
None: 0
Query
XK?
Example
Set the start trigger to none.
XK 0
When setting the start trigger to external input
Syntax
XK p1<terminator>
p1
Start trigger type
External input:
1
Query
XK?
Example
Set the start trigger to external input.
XK 1
When setting the start trigger to the input level
Syntax
XK p1,p2,p3,p4,p5
<terminator>
p1
Start trigger type
Analog input level:
2
p2
Channel to be used as the start trigger
For 8 ch input:
01
to
08
For 16 ch input:
01
to
16
Pulse CH:
PL01
Calculation CH:
CA01 to CA32
Communication CH:
CO01 to CO32
p3
Start trigger level type
High limit trigger:
H
Low limit trigger:
L
Window IN:
I
Window OUT:
O
p4
High limit value
-99999
to
99999
p5
Low limit value
-99999
to
99999
Query
XK?
Example
Set the start trigger to Window IN of analog input
CH3, and set the high and low limits to 2000 and
1000, respectively.
XK 2,03,I,2000,1000
Description Set the high and low limits using a five-digit
integer without a decimal point. The decimal
place is the same as the span and scale settings
of the analog input channel.
Be sure to set the high limit greater than the low
limit.
When setting the start trigger to an alarm
Syntax
XK p1,p2<terminator>
p1
Start trigger type
Alarm:
3
p2
Alarm output number to be used as the start
trigger
1
to
4
Query
XK?
Example
Set the start trigger to alarm output number 4.
XK 3,4
When setting the start trigger to a specific time
Syntax
XK p1,p2,p3
p1
Start trigger type
Time:
4
p2
Start date (YY/MM/DD fixed form)
YY (year):
00
to
99
MM (month):
01
to
12
DD (day):
01
to
31
p3
Start time (HH/MM/SS fixed form)
HH (hour):
00
to
23
MM (min):
00
to
59
SS (s):
00
to
59
Query
XK?
Example
Set the start trigger to 3:45:6 on January 2, 2006.
XK 4,06/01/02,03:45:06
5.7 Basic Setting Commands