3-3
IM XL100C-E
1
2
3
4
5
6
7
App
Index
Serial Interface
Precautions Regarding Data Reception Control
When handshaking is used to control the reception of data, data may still be sent from
the PC even if the free space in the receive buffer drops below 256 bytes. In this case,
after the receive buffer becomes full, the excess data will be lost, whether or not
handshaking is in effect. Data storage of data resumes when there is free space in the
buffer.
Setting the RS-232 Interface
• Selecting the Slave Address
Select the address from the following values.
1 to 32
• Selecting the Baud Rate
Select the baud rate from the following:
2400, 4800, 9600, 19200, or 38400
• Setting the Data Length
Select the data length from below. To output data in binary format, be sure to set the
data length to 8 bits.
7 or 8
• Setting the Stop Bit
Select the stop bit from the following:
1 or 2
• Selecting the Parity Check
Select the parity check from the following:
ODD, EVEN, or NONE
• Selecting the handshaking
Select the handshaking method from the following.
OFF, XON/XOFF, or CS/RS
Procedure
1.
Press
SETTING
to enter Setting Mode.
2.
Use the arrow keys to select COMMUNICATION, and press SELECT.
→
3.1 RS-232 Interface Specifications and Setup Procedure