< 9. Troubleshooting >
9-1
IM 12F5A1-01E
9. TROUBLESHOOTING
9.1 When <failure> is Indicated
Daily maintenance for maintaining good FC400G operation is described in Chapter 8; this chapter
sets forth how to maintain the FC400G in the event of error.
9.1.1 Indication of Error Message or Error Code
Types of <failure> and respective error messages / error codes are listed in Table 9.1.
• Errors occurring during Measurement
If error occurs during measurement, the LCD displays an error message (Err. ) and
FAIL
.
The FC400G also generates fail contact output (fail LED lights).
• Errors occurring during Maintenance
If error occurs during maintenance (during zero and / or span calibration, or during plateau
characteristic collection), the LCD displays an error message (Err. ) and
FAIL
to inform the
operator of error signal occurrence. However, no fail contact output is generated and output
signal overshooting does not function.
• Error occurring during Data Setting
If data set during data setting is incorrect, an error message (NOTE.) appears for approx. 2
sec to inform the operator of incorrect data setting. In this case, set data does not update the
previous; set correct data again.
Table 9.1
Error Message List
Error No.
Error Content
Classification
Err. 1
Err. 2
Diffusion current value exceeded
Sample temperature error
Occurs during measurement or maintenance.
(Err. 3)
Reserved error code (presently not used)
Err. 4
Applied voltage error
Occurs during measurement or maintenance.
Err. 5
Err. 6
Err. 7
Slope error
Zero point error
Response error during zero or span
calibration
Occurs during maintenance (however, only
during zero or span calibration).
Err. 8 (Note 1)
Temperature compensation range
exceeded
Occurs during measurement or maintenance.
(Err. 9)
Reserved error code (presently not used)
——————
Err. 10
Converter failure
Occurs during measurement or maintenance.
NOTE. 01
NOTE. 02
NOTE. 03
Input data setting range exceeded
Input data setting error
Input data setting error
Occurs during data setting.
Note 1: This error Err.8 occurs only if exceeded temperature compensation range check is set to ON in service code 02.
9.1.2 Remedies When Errors Occurs
The possible cause of each error message is described in the following.
Take appropriate steps for each.