< 7. Calibration >
7-5
IM 12F5A1-01E
• Power is ON and zero filter flushing start conditions are satisfied.
• Auto zero calibration interval is at least one day and Auto calibration functions are ON.
• When Auto calibration functions are ON and Auto zero calibration interval is changed to at least
one day.
• When Auto zero calibration is ON and Auto zero calibration interval is at least one day, and
“Wash / Auto zero calibration setting” is switched to Auto zero calibration.
When timer (are) starts, counter is reset.
Counter is stopped when:
• Zero filter flushing timer is ON and Auto calibration setting is turned OFF.
• Zero filter flushing timer is ON and Auto zero calibration interval is reduced to less than one day.
• Zero filter flushing timer is ON and Wash / Auto zero calibration setting is changed to Wash.
NOTE
1) During zero filter flushing, if Auto zero calibration is started then zero filter flushing is aborted.
2) If the “zero filter flushing counter” reaches its “count up” setting while the instrument is in auto
zero calibration mode, zero filter flushing is cancelled.
3) During zero filter flushing, only
MODE
key and
*
key are enabled pressing either of these
keys aborts zero filter flushing, and immediately the analog output hold is released, <HOLD>
and <FLUSH> go OFF, and the <MAINT> contact output goes OFF.
4) Counting continues regardless of timer operation mode.
5) In <MAINT> mode, if the “zero filter flushing counter” reaches its “count up” setting then zero
filter flushing is cancelled - not triggered - that time: the timer is reset and counting restarts.