<7. Startup>
7-17
IM 11M12A01-04E
(3) Oxygen concentration in span gas
Enter the span gas oxygen concentration for calibration. If instrument air is used, enter 21
vol % O
2
. When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a
span gas), use a hand-held oxygen analyzer to measure the actual oxygen concentration,
and then enter it.
CAUTION
If instrument air is used for the span gas, dehumidify the air to a dew point of -20°C and remove
any oil mist or dust.
Incomplete dehumidifying or unclean air will have an adverse effect on the measurement
accuracy.
Table 7.14
Calibration Parameter Codes
Set item
Parameter code
Set value
Calibration mode
B03
0 Manual calibration
1 Semi-automatic calibration
2 Automatic calibration
Zero gas oxygen concentration
B01
Enter oxygen concentration.
Span gas oxygen concentration
B02
Enter oxygen concentration.
11th Edition : Jul. 19, 2017-00