<5. Wiring>
5-6
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
5.5.1
Cable Specifications
Use a 2-core or 3-core cable for this wiring. Depending on the number of input(s), determine
which cable to use.
5.5.2
Wiring Procedure
(1) M4 screws are used for the terminal of the converter. Each cable should be equipped with
the corresponding crimp contact.
(2) The ON/OFF level of this contact input is identified by the resistance. Connect a contact
input that satisfies the descriptions in Table 5.2.
Table 5.2
Identification of Contact Input ON/OFF
Closed
Open
Resistance
200 Ω or less
100 kΩ or more