<8. Detailed Data Setting>
8-20
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
n
Default Values
When the analyzer is delivered or data are initialized, the parameters are by default as shown in
Table 8.20.
Table 8.20
Parameter Codes for Exhaust Gas Temperature and Pressure Settings
Item
Default setting
Exhaust gas temperature
300°C
Exhaust gas pressure
101.33 kPa abs.
8.6.5
Setting Purging
Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration
gas for a given length of time before warm-up of the detector. This prevents cell breakage during
calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge
time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100°C or below upon power up and the purge
time is set in the range of 1 to 60 minutes.
F8.5.ai
Displayed alternately
Figure 8.4
Display during Purging
n
Procedure
Use the parameter-code table below to set the purging time.
The allowable input ranges from 0 to 60 minutes.
Table 8.21
Purging Time
Set item
Parameter code Set range
Units
Purging time
F15
0 to 60
minutes
n
Default Value
When the analyzer is delivered, or if data are initialized, purging time is set to 0 minutes.