<7. Startup>
7-3
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
2: Infrared switch
4: Decimal point
1: Data display area
3: Engineering-unit
display area
>
>
ENT
µ M m N k g a l b b l
%
scftm3 /d /s /h /m
F7.3E.ai
Figure 7.4
Infrared switch and the display
1. Data display area: Displays the oxygen concentration, humidity, set values, alarm numbers,
and error numbers.
2. Infrared switch: Three switches perform data setting operations.
3. Engineering-unit display area: the percent sign appears when the oxygen concentration or
humidity is displayed.
4. Decimal point:
A decimal point is displayed.
Table 7.1
Switch and Function
Switch
Function
>
1.
Moves the position of the digit to the right. If you continuously touch the key, the position
of the digit will move continuously to the right, finally returning to the leftmost position
after reaching the rightmost position of the digit.
2.
Selects Yes or No.
3.
When you touch this key together with the [ENT] key, the previous display then appears,
or the operation will be cancelled.
^
Used to change values. If you continuously touch this key, the value of the digit will increase
continuously, e.g., from 1 to 2 to 3 (for numeric data), or from A to B to C (for alphabetic
characters), and finally return to its original value.
ENT
1. Used to change the basic panel display to the parameter selection display.
2. Used to enter data.
3. Advances the operation.
The three infrared switches are activated by completely touching the glass surface of the switch.
To touch any of the keys continuously, first touch the surface and then completely remove your
finger from the surface. Then touch it again.
Infrared switches consist of two elements: an infrared emitting element and an infrared
acceptance element. Infrared light-waves from the element bounces on the operator’s finger
and are reflected back to the acceptance element, thereby causing the infrared switch to turn on
and off, depending on the strength of the reflected light-waves. From this operating principles,
carefully observe the following:
CAUTION
1. Be sure to put the equipment case cover back on. If this is not done, the infrared switch will
not reflect the infrared light-waves, and a “dSPErr” error will be issued.
2. Before placing the equipment in operation, be sure to wipe off any moisture or dust on
the glass surface if it is wet or dirty. Also make sure your fingers are clean and dry before
touching the glass surface of the switch.
3. If the infrared switches are exposed to direct sunlight, they may not operate correctly. In
such a case, change position of the display or install a sun cover.