5-16 Parameter setting
IM 12J6B3-E-H
5-3-2. Temperature functions
Code 10
*T.SENS
Select temperature sensor to suit the measuring probe. Refer to instructions with
DO sensor for which temperature sensor is used.
Code 11
*T.UNIT
Select °C or °F for temperature display.
Code 12
*T.ADJ.
Adjust offset in temperature measurement when the sensor is at a stable known
value.
Code 13
*T.MAN.
Enable and set manual temperature compensation.
The maximum and minimum slope values that may be entered depend on the nominal current from service
code 01.
Example
: When nominal current = 3.75 µA the maximum and minimum slope calibration values that
may be entered are:
Maximum
3.75 x 2 = 7.5 µA
Minimum
3.75 x 0.4 = 1.5 µA