10-4 Appendix
IM 12J6B3-E-H
10-6. Software Revision History
Software changes of the DO402
10-6-1. Changes made by software release 1.1
●
Slope calculation bug is solved in release 1.1. In release 1.0 calibration caused a wrong slope calculation,
resulting in a wrong measurement.
●
The Air calibration menu in the maintenance mode is expanded with a zero current (0%) calibration, used
for calibration in nitrogen.
●
When the manual temperature in service code 13 is enabled then a manual temperature adjust menu is
accessible in maintenance mode for quick adjustment of the temperature.
●
Software changes for better performance with polarographic sensors.
10-6-2. Changes made by software release 1.2
●
Arrow in display in case of AIR.CAL was not correct. This has been changed
●
Maintenance timer was decrementing in case of a powerdown-powerup sequence.
●
Slope value was not reset to 100% in case of Defaults load. (SV70)
●
Init userflow was not correct. (procedure was aborted in case of 30 sec. of key-silence)
●
Communication changes to operate with the DO402-PC package.
10-6-3. Changes made by software release 1.3
●
The maximum wash interval was 18.2 hours instead of the specified 36 hours
●
When the slope was manually changed using service code 22, the change had no effect on the process
value.
●
The instrument showed negative values as 0.0 ppm, mg/l or %sat, therefore a negative value (wrong air
calibration) was never showed to the user.
●
The sensor initialization method was not very user friendly.
●
The temperature measurement with the Ingold compatible sensor was incorrect.
10-6-4. Changes made by software release 1.4
●
In S.C. 51 manual pressure can be enabled. In maintenance mode the pressure can be entered. If manual
pressure is enabled, a warning is given when starting H
2
O, manual or air calibration.
●
As a kind of remote range a range 1B is added to the already existing range 1 and range 2. Service Code
38 can be used to allow range 1B. (and automatically disables WASH with external button).
10-6-5. Changes made by software release 1.5
●
Default no membrane check.
●
Stepping through Service Code 01 no longer defaults zero and slope.
●
Interpolation mA-table improved.
●
2-point calibration improved.
●
FAIL contact no longer irregularly toggles.
10-6-6. Changes made by software release 1.6
●
Bug fixed in configuration contact S4 as control contact.