![YOKOGAWA EJA565E User Manual Download Page 26](http://html1.mh-extra.com/html/yokogawa/eja565e/eja565e_user-manual_3759832026.webp)
26
IM-P-20200627-01
7.
Programming
7.1 Programming with Hand Held Terminal
CAUTION
•
When using HART or a Hand Held Terminal (HHT), a minimum resistance of
250 Ω
must be present in
the loop of the 2-wire system. This is necessary for proper communication (see drawing below). A power
supply of at least
17 Vdc
must be used.
The EJA565E can easily be programmed with the Hand Held Terminal (
YHC
HART Communicator).
Option 1: HART Hand
H
eld terminal connected across the transmitter.
Option 2: HART Hand
H
eld terminal connected across the loop resistor.
−
+
Power
Supply
250 Ω
17 - 36 Vdc
−
+
Power
Supply
250 Ω
17 - 36 Vdc
<7. Explanation
-
Programming Points>