IM 01C25C01-01E
9-14
9. GENERAL SPECIFICATIONS
SUPPLY
CHECK
or
ALARM
+
–
+
–
●
Terminal Configuration
Power supply and output terminal
External indicator (ammeter) terminal
*1*2
or
Status contact output terminal
*2
(when /AL is specified)
Ground terminal
Communication
terminals (BT200 etc.)
connection hook
●
Terminal Wiring
*1: When using an external indicator or check meter, the internal
resistance must be 10
or less. A check meter or indicator
cannot be connected when /AL option is specified.
*2: Not available for fieldbus communication type.
Check meter
connection hook
*1*2
SUPPLY –
CHECK – or ALARM –
CHECK + or
ALARM +
+
–
F0911.EPS
< Factory Setting >
Tag Number
Amplifier
Damping
*2
As specified in order
‘2 sec.’ or as specified in order
Display Setting
As specified in order
'0 to 25MPa' for M and H capsule.
Display unit is selectable from the units
listed in above 'Calibration Range Units'.
Calibration Range
Units
Designated differential pressure value
specified in order. (%, Engineering unit or
user scaled value.) Display mode 'Linear'
or 'Square root' is also as specified in
order.
As specified in order
Selected from mmH
2
O, mmH
2
O(68˚F),
mmAq, mmWG, mmHg, Pa, hPa, kPa,
MPa, mbar, bar, gf/cm
2
, kgf/cm
2
, inH
2
O,
inH
2
O(68˚F),inHg, ftH
2
O, ftH
2
O(68˚F) or
psi. (Only one unit can be specified)
Calibration Range
Upper Range Value
Calibration Range
Lower Range Value
T0914.EPS
Static Pressure
Display Range
*1
*1: For differential pressure transmitters only.
*2: To specify these items at factory, /CA or /CB option is required.