![YOKOGAWA DpharpEJX vigilantplant EJX110 User Manual Download Page 57](http://html1.mh-extra.com/html/yokogawa/dpharpejx-vigilantplant-ejx110/dpharpejx-vigilantplant-ejx110_user-manual_3738051057.webp)
IM 01C25B01-01E
9-12
9. GENERAL SPECIFICATIONS
Item
Description
Code
Material certificate
*1
Cover flange
*2
Cover flange, Process connector
*3
Test Pressure: 16 MPa(2300 psi)
*4
Test Pressure: 25 MPa(3600 psi)
*5
Test Pressure: 3.5 MPa(500 psi)
*6
Test Pressure: 500 kPa(2000 inH
2
O)
*7
Test Pressure: 50 kPa(200 inH
2
O)
*8
Test Pressure: 32 MPa(4500 psi)
*9
Test Pressure: 50 MPa(7200 psi)
*10
M01
M11
T12
T13
T01
T11
T04
T09
T08
Pressure test/
Leak test certificate
Nitrogen(N
2
) Gas
*11
Retention time: one minute
T0914.EPS
Nitrogen(N
2
) Gas or water
*12
Retention time: one minute
*1: Material traceability certification, per EN 10204 3.1B.
*2: Applicable for Process connections code 0 and 5.
*3: Applicable for Process connections code 1, 2, 3, and 4.
*4: Applicable for Capsule code L of EJX110A and Capsule code B of EJX430A and EJX310A.
*5: Applicable for Capsule code M and H of EJX110A.
*6: Applicable for Capsule code A of EJX430A and EJX310A.
*7: Applicable for Capsule code H of EJX430A.
*8: Applicable for Capsule code M of EJX310A.
*9: Applicable for EJX130A and Capsule code C of EJX440A.
*10: Applicable for Capsule code D of EJX440A.
*11: Pure nitrogen gas is used for oil-prohibited use (Option code K1, K2, K5, and K6).
*12: Pure nitrogen gas or pure water is used for oil-prohibited use (Option code K1, K2, K5, and K6).