6-2
IM AQ6376E-02EN
Item
Specifications
Safety standards
EN 61010-1, EN60825-1
Overvoltage category II
10
Pollution degree 2
11
Laser
EN 60825-1:2014, IEC 60825-1:2007, GB 7247.1-2012 class 1
Emissions Compliant standards
EN 61326-1:2013 Class A Group 1 Table 2
EN 61000-3-2
EN 61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand EN 61326-
1:2013 Class A Group 1 Table 2
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
This product is classified as Class A (for use in industrial environments).
Operation of this product in a residential area may cause radio interference,
in which case the user will be required to correct the interference.
Cable conditions
• Ethernet connector
Use category 5 or better Ethernet cables.
12
• VIDEO OUT connector
Use a shielded D-sub 15 pin VGA cable.
12
• USB port
Use USB peripherals (e.g., mouse) with shielded USB cables.
12
• GP-IB interface connector
Use a shielded GP-IB cable.
12
• TRIGGER IN, TRIGGER OUT, ANALOG OUT terminal
Use a BNC cable.
12
Immunity Compliant standards
EN 61326-1 Table 2
Influence in the immunity environment (criteria A)
Wavelength measurement accuracy: ±0.1 nm
Cable conditions
The same as the cable conditions listed above for emissions
Environmental standards
13
EU RoHS Directive compliant
1 Horizontal axis scale: In wavelength display mode
2 9.5/125 μm single mode fiber, after warm-up of 2 hours, after alignment adjustment with built-in wavelength reference light
source
3 Vertical scale: absolute value level display mode, resolution setting: 0.1 nm or more.
4 When using 9.5/125 μm single mode fiber (SSMA type in JIS C6835, PC polishing, mode field diameter: 9.5 μm, NA: 0.104 to
0.107)
5 After wavelength calibration with built-in reference light source, sampling interval: 0.003 nm or less, sensitivity: MID, HIGH1,
HIGH2, or HIGH3
6 Pulse light measurement mode: OFF
7 Span 100 nm or less, wavelength sampling points: 1001, averaging times: 1
8 Certain commands may not support the AQ6317 depending on the relationship between the target model specifications and
functions.
9 The LDC display may contain defective pixels (always ON or always OFF). (0.002% or fewer of all pixels including RGB).
Does not indicate a general malfunction.
10 The overvoltage category is a value used to define the transient overvoltage condition and includes the rated impulse
withstand voltage. Category II applies to electrical equipment that is powered through a fixed installation, such as a wall outlet
wired to a distribution board.
11 Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas that deteriorates withstand voltage or surface
resistivity. Pollution Degree 1 applies to sealed space (with no pollution or only dry non-conductive pollution). Pollution Degree
2 applies to normal indoor atmospheres (with only non-conductive pollution).
12 Use a cable with a length of 3 m or less.
13 For conformity to environmental regulations and/or standards other than EU, contact your nearest YOKOGAWA office