36
IM AQ6375-02EN
Item
Specifications
Function
Marker/Search
Delta markers (1024 points maximum), vertical/horizontal line markers, peak search, 2nd
peak search, bottom search, 2nd bottom search, auto search (ON/OFF), search between
vertical axis line markers, search within zoom area
Analysis
Spectral width analysis (threshold, envelope, RMS, Peak RMS, notch), WDM (OSNR)
analysis, EDFA-NF analysis, filter peak/bottom analysis, WDM filter peak/bottom analysis,
DFB-LD analysis, FP-LD analysis, LED analysis, SMSR analysis, power analysis, PMD
analysis, Pass/Fail judgment from template, auto analysis, analysis between vertical axis
line markers, analysis within the zoom area
Other
Auto alignment function with built-in reference light source and automatic wavelength
calibration function
Data storage
Internal memory
64 traces, 64 programs, 3 templates
Internal storage
128 MB max
External storage
USB storage media (USB memory/HDD), format: FAT32
File types
CSV (text), binary, bitmap, TIFF
Interfaces
Remote control
GP-IB, RS-232, Ethernet (TCP/IP)
AQ6317 series compliant commands (IEEE488.1) and IEEE488.2
Categories
GP-IB x 2 (for standard and external control), RS-232, Ethernet, USB1.1 x 2, PS/2 (for
keyboard), SVGA output, analog output port, trigger input port, trigger output port
Optical connectors
For optional I/O, AQ9447(*) connector adapter (option) required.
For wavelength reference light source output, AQ9441(*) universal adapter (optional) required.
(*): Connector types: FC, SC, ST
Printer
Built-in thermal printer (factory option)
Display
12
10.4" color LCD (resolution: 800 x 600 pixels)
Power requirement
100 to 240 VAC, 50/60 Hz, approximately 150 VA
Environment conditions
Operating temperature range: +5 to +35°C
Operating humidity range: -10 to +50°C
Ambient humidity: 80% RH or less (no condensation)
Recommended calibration period
1 year
External dimensions
11
Approximately 426 (W) x 221 (H) x 459 (D) mm
Mass
Approximately 27 kg (excluding built-in printer)
Safety standards
Conforming standards
EN61010-1
EN60825-1
Pollution degree 2
13
Emissions
Conforming standards
EN61326 Class A
EN55011 Class A, Group 1
EN61000-3-2
EN61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand
EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard
(
한국 전자파적합성기준
)
This is a class A instrument (industrial use). Wireless
interference may occur in home environments. If so, the user
must take appropriate countermeasures.
Cable conditions
• TRIGGER IN, TRIGGER OUT, ANALOG OUT terminal.
Use a BNC cable
14
• Use a serial (RS-232) interface connectorand RS-232
shielded cable.
14
• Use an Ethernet connector and a category 5 or higher
Ethernet cable.
15
• Use a VIDEO OUT connector and a D-sub 15pin VGA
shielded cable
14
• Use a USB peripheral (such as a mouse) that uses a USB
port and shielded cable
14
• Use a keyboard connector and PS/2 shielded cable
14
• Use the GP-IB1 or GP-IB2 interface connector and a GP-
IB shielded cable
14
10 Specifications