35
IM AQ6375-02EN
10
Specifications
Item
Specifications
Applicable fiber
SM (9.5/125
µ
m), GI (50/125
µ
m, 62.5/125
µ
m)
Measurement wavelength range
1
1200 to 2400 nm
Span
1
0.5 nm to 1200 nm (full wavelength range), 0 nm
Wavelength accuracy
1, 2, 3
±0.05 nm (1520 to 1580 nm)
±0.1 nm (1580 to 1620 nm)
±0.5 nm (full wavelength range)
Wavelength repeatability
1, 2
±0.015 nm (1 minute)
Wavelength sampling point
101 to 50001, AUTO
Wavelength resolution setting
1, 2
0.05, 0.1, 0.2, 0.5, 1.0, 2.0 nm
Level sensitivity setting
10
NORM_HOLD, NORM_AUTO, NORMAL, MID, HIGH1, HIGH2 and HIGH3
Level sensitivity
2, 4, 5, 7
-62 dBm (1300 to 1500 nm, resolution: 0.1 nm or more, measuring sensitivity: HIGH3)
-67 dBm (1500 to 1800 nm and 2200 to 2400nm, resolution: 0.1 nm or more, measurement
sensitivity: HIGH3)
-70 dBm (1800 to 2200 nm, resolution: 0.1 nm or more, measurement sensitivity:HIGH3)
Level accuracy
2, 4, 5, 6
±1.0 dB (1550 nm, input level: -20 dBm, measuring sensitivity: MID, HIGH1, HIGH2,
HIGH3)
Leval linearity
2, 4
±0.05 dB (input level: -30 to +10 dBm, measuring sensitivity: HIGH1, HIGH2, HIGH3)
Maximum input power
2, 4
+20 dBm (per channel, full span)
Safe max. input power
2, 4
+25 dBm (total light input power)
Close-in dynamic range
1, 2, 9
45 dB (1523 nm, ±0.4 nm of peak wavelength, resolution: 0.05 nm)
55 dB (1523 nm, ±0.8 nm of peak wavelength resolution: 0.05 nm)
Polarization dependency
2, 4, 6
±0.1 dB (1550 nm)
Sweep time
1, 7, 8
0.5 s (NORM_AUTO), 1 s (NORMAL)
10 s (MID)
20 s (HIGH1), 60 s (HIGH2), 600 s (HIGH3)
Function
Automatic measurement
Program function (64 programs, 200 steps)
Setting of measuring conditions
Center wavelength, span, wavelength sampling points, wavelength resolution,
measurement sensitivity, averaging times (1 to 999), sweep (single, repeat, AUTO:
automatically sets measuring conditions), sweep between marker function, pulse light
measurement function, external trigger measurement function, sweep status output
function, analog output function, synchronous sweep with turnable laser source function,
air/vacuum wavelength measurement function, template-based Pass/Fail judgment function
Display
Level scale (0.1 to 10 dB/div., linear scale), level subscale (0.1 to 10 dB/div., linear scale),
reference level display, vertical axis DIV display (8, 10, 12), horizontal axis wavelength
(nm)/frequency (THz)/wavenumber (cm
-1
) display, horizontal axis scale zoom in/out display,
measuring conditions display, noise mask display, data table display, label display, split
screen display, percent display, power density (dB/nm) display, dB/km display, template
display
Traces
Simultaneous display of 7 independent traces, write mode fixed mode setting, show/hide
setting, max/min value detection display, calculation between traces display, roll averaging
display (sweep average) (2 to 100 times), normalized display, curve fit display (peak curve
fit, marker curve fit), trace copy function, trace clear function, saving/loading all trace
function