Well Ventilated Location
Ventilation holes are present at the sides and rear of the instrument. To keep the
internal temperature from rising, always maintain a gap of 200 mm or more between the
ventilation holes and the installation surfaces.
Also be sure to maintain sufficient clearance for connecting measurement cables.
There are inlet holes on the side and
bottom of the instrument.
20 cm or more
20 cm or
more
20 cm or
more
20 cm or
more
Ambient Temperature and Humidity
Ambient temperature: 5–35°C
Ambient humidity: 80% RH or lower (no condensation present)
Note
Condensation may occur if the instrument is moved to another place where the ambient
temperature is higher, or if the temperature changes rapidly. In this case, let the instrument
adjust to the ambient temperature for at least one hour before use.
When the instrument is packaged in a box and moved, prevent condensation by allowing
sufficient time for the instrument to acclimatize before removing it from the box.
Flat, Even Surface
Install the instrument with the correct orientation on a stable, horizontal surface. If the
instrument is installed in a horizontal position, rubber stoppers can be attached to the
feet at the rear of the instrument to prevent the instrument from sliding. One set of rubber
stoppers (two stoppers) are included with the instrument.
Rubber stoppers
A9088ZM
Feet at the rear of the instrument
Rack Mounting
To rack-mount the instrument, use the separately sold rack mount kit.
Product Name
Model
Rack Mounting Kit for EIA Single
751535-E5
Rack Mounting Kit for JIS Single
751535-E5