42
IM AQ6374-02EN
Emissions
Conforming
standards
EN61326-1 Class
A
EN55011
Class
A,
Group
1
EN61000-3-2
EN61000-3-3
EMC standards of Australia and New Zealand EN55011 Class A, Group 1
Korea Electromagnetic Conformity Standard
(
한국 전자파적합성기준
)
This is a class A instrument (industrial use). Wireless interference
may occur in home environments. If so, the user must take
appropriate countermeasures.
Cable conditions
• TRIGGER IN, TRIGGER OUT, ANALOG OUT terminal.
Use a BNC cable
11
• Use a serial (RS-232) interface connectorand RS-232 shielded
cable.
11
• Use an Ethernet connector and a category 5 or higher Ethernet
cable.
11
• Use a VIDEO OUT connector and a D-sub 15pin VGA shielded
cable
11
• Use a USB peripheral (such as a mouse) that uses a USB port
and shielded cable
11
• Use the GP-IB interface connector and a GP-IB shielded
cable
11
Immunity
Conforming standards
EN61326-1 Table 2 (For use in industrial locations)
Effect in immunity environment
Wavelength measurement sensitivity:
Within ±0.1 nm
Cable conditions
Same as above emission cable conditions.
1: Horizontal axis scale: In wavelength display mode
2: 9.5/125
μ
m single mode fiber, after alignment adjustment with a built-in wavelength reference light source, when the purge
gas is not used.
3: Vertical scale: absolute value level display mode, resolution setting: 0.2 nm or more.
4: When using 9.5/125
μ
m single mode fiber (SSMA type in JIS C6835, PC polishing, mode field diameter: 9.5
μ
m, NA: 0.104 to
0.107)
5: Resolution: 0.05 nm
6: Pulse light measurement mode: OFF
7: Span 100 nm or less (excluding measurement wavelength range 570 nm to 580 nm and 900 nm to 1000 nm), wavelength
sampling points: 1001, averaging times: 1
8: Certain commands may not support the AQ6317 depending on the relationship between the target model specifications and
functions.
9: The LDC display may contain defective pixels (always ON or always OFF).
(0.002% or fewer of all pixels including RGB). Does not indicate a general malfunction.
10: Pollution degree refers to the degree of adherence by a solid, liquid, or vapor that reduces the withstand voltage or surface
resistance factor. Pollution degree 1 applies to closed atmospheres (no pollution, or only dry, non-conductive pollution).
Pollution degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
11: Use a cable of 3 m in length or less.
10 Specifications