2.2 Installing the Instrument
WARNING
• This instrument is designed to be used indoors. Do not install or use it outdoors.
• Install the instrument so that you can immediately remove the power cord if an abnormal or
dangerous condition occurs.
CAUTION
• If you block the inlet holes on the bottom or the exhaust holes on the rear side of the
instrument, the instrument will become hot and may break down. Also, high temperatures
may shorten the service life of the internal reference light source (He-Ne laser).
• Do not place this instrument on top of another device that is emitting heat. If heat enters
through the inlet holes on the bottom of the instrument, the instrument will become hot and
may break down.
French
AVERTISSEMENT
• L’instrument est prévu pour une utilisation en intérieur. Ne pas l’installer, ni l’utiliser à
l’extérieur.
• Installer l’instrument de manière à pourvoir immédiatement le débrancher du secteur en
cas de fonctionnement anormal ou dangereux.
ATTENTION
• Si vous bloquez les orifices d’entrée situés sous l’instrument ou les orifices de sortie
arrière, l’instrument surchauffe et risque de tomber en panne. En outre, les températures
élevées risquent d’écourter la durée de vie de la source lumineuse de référence interne
(laser He-Ne).
• Ne pas placer cet instrument au dessus d’un autre dispositif qui émet de la chaleur. Si
la chaleur pénètre à travers les orifices d’entrée situés au sous l’instrument, celui-ci se
chauffera et pourrait tomber en panne.