18-5
IM AQ1200-01EN
Troubleshooting, Maintenance, and Inspection
18.2 Error Messages
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Index
App
Code
Message
911
The Voltage of AC power supply is too low. The instrument may be damaged if it is kept using in this condition. The
instrument will be powered off in 10 sec. Please make sure of using the dedicated AC adapter.
912
The Voltage of AC power supply is too high. The instrument may be damaged if it is kept using in this condition. The
instrument will be powered off in 10 sec. Please make sure of using the dedicated AC adapter.
913
Battery is low. The instrument may be damaged if it is kept using in this condition. The instrument will be powered off
in 10 sec. Please use AC adapter.
914
Errors in the battery or in the charging circuit. The instrument may be damaged if it is kept using in this condition.
The instrument will be powered off in 10 sec. Please remove the battery and use AC adapter.
915
Battery temperature is too high. Please stop measuring, and wait for a while.
916
The temperature inside the instrument is increasing. Please stop measuring, and wait for a while.
917
The temperature inside the instrument is increasing. Please stop measuring, and wait for a while.
918
The temperature inside the instrument is too high. The measurement was aborted. Please power it off.
921
There is incoming light. Or, optical module may be damaged.
922
Incorrect date and time setting. Set the correct date and time.
923
The ambient temperature is out of range. The light source cannot be turned on.
925
Please use AC adapter.