Detailed Setup Screen
Press the
SETUP
soft key and then the Data Save Logging soft key to display the following
screen.
Set the measurement interval (500ms, 1s, 2s, 5s, 10s).
Set the logging count (10 to 36000).
The time to perform logging measurements over is
calculated automatically from the measurement
interval and the logging count and displayed.
File Save Screen
The path of the destination folder
The file name candidate for the next save operation
The number of folders and files in the destination folder
Set the file name (see section 16.2).
Saves the data
Set the destination drive
(internal memory, USB memory).
Explanation
Creating and Saving Logs
This instrument is always measuring power when the Power Meter screen is displayed. To save the
measured values, you must log them.
• To start logging, press the Logging START soft key. During logging, on the menu, “Logging START”
changes to “Logging STOP.”
• Logging stops when the specified number of logs are recorded or when you press the Logging
STOP soft key. Then, on the menu, “Logging STOP” changes to “Logging START,” and the file save
screen appears.
• You can save logged results in CSV format.
Measurement Interval
You can set the interval for logging measured values to one of the following values.
500 ms, 1 s, 2 s, 5 s, 10 s
Logging Count
You can set the number of values to log within the following range.
10 to 36000
3.3 Logging Measured Values and Saving Logged Results