IM 01E20F02-01E
8-5
8. PARAMETER LISTS
T0802-2.EPS
Write
Mode
Explanation
Factory
Default
Relative
Index
Parameter Name
Index
The linearization type of sensor output. AXF is "linear with input"
N/A. For the future use.
N/A. For the future use.
This parameter is used to select the language for use on the
display. Now only 1: English is available.
This parameter is used in order to set the LCD display mode.
1: Flow Rate (%) 2: Flow Rate 3: Integrator1 Out
4: Integrator2 Out 5: Arithmetic Out can be selected.
This parameter is used in order to set the LCD display mode.
1: Off 2: Flow Rate (%) 3: Flow Rate 4: Flow Rate (Bar)
5: Integrator1 Out 6: Integrator2 Out 7: Arithmetic Out 8: PD Tag
9: Adhesion Check 10: Communication can be selected.
This parameter is used in order to set the LCD display mode.
1: Off 2: Flow Rate (%) 3: Flow Rate 4: Flow Rate (Bar)
5: Integrator1 Out 6: Integrator2 Out 7: Arithmetic Out 8: PD Tag
9: Adhesion Check 10: Communication can be selected.
This parameter is used in order to select the unit of size
(diameter) of the sensor (flowtube).
This parameter is used in order to set the size (diameter) of the
sensor (flowtube). If the set value exceeds the valid range, the
warning "72:Size Set Err" will be displayed.
Sets the time constant of damping for the flow rate calculation.
This parameter execute the automatic zero adjustment function:
If "2:Execute" is selected, this function will be started. "Now Auto
Zero Executing..." is indicated whle the Auto Zero function being
carried out and after finishing the adjustment, this parameter is
set to "1:Cancel". The result of the automatic zero adjustment is
confirmed using MAGFLO_ZERO, and if the result exceeds the
rated value, the warning"82:Auto Zero Wng" will be displayed.
This parameter shows the current zero point compensation value
for the sensor. This parameter is used to display the results
obtained from AUTO_ZERO_EXE. Specifically, the correction
values displayed, and it is also possible to directly enter correction
values. This parameter must not be downloaded by the operator
This parameter is used in order to set the low-frequency meter
factor.
This parameter is used in order to set the high-frequency meter
factor.
This parameter is used in order to set the low-frequency meter
factor as required when Enhanced DF (i.e., enhanced dual
frequency excitation) is selected. If "Standard DF" has been
selected for MEASURE_MODE, neither LOW_MF (EDF) nor
HIGH_MF (EDF) is displayed.
This parameter is used in order to set the high-frequency meter
factor as required when Enhanced DF (i.e., enhanced dual
frequency excitation) is selected.
Always 1:ADMAG AXF
Selects measurement mode for dual frequency.
Selectable 1: Standard DF 2: Enhanced DF.
Set the low cut range corresponding 0 to 10% of the range.
Set the mode of the flow measurement, either unidirectional or
bidirectional.
28 2028
30 2030
LANGUAGE
1:
English
Auto
31
2031 DISPLAY_SELECT1
2: Flow Rate
Auto
32 2032
DISPLAY_SELECT2
1:
Off
Auto
33 2033
DISPLAY_SELECT3
1:
Off
Auto
34 2034
1013:
mm
O/S
35 2035
NOMINAL_SIZE
100
O/S
36 2036
3.0
O/S
37 2037
AUTO_ZERO_EXE
1:
Cancel
Auto
40 2040
HIGH_MF
1.0000
O/S
41 2041
LOW_MF(EDF)
1.0000
O/S
42 2042
HIGH_MF(EDF)
1.0000
O/S
43
2043
1: ADMAG AXF
O/S
44
2044 MEASURE_MODE
1: Standard DF
O/S
45 2045
0
O/S
46 2046
BI_DIRECTION
1:
Unidirectional O/S
Assign an arbitrary positive or negative sign to the measured
PV value.
1. Positive
2: Negative
47 2047
FLOW_DIRECTION 1:
Positive
O/S
38 2038
MAGFLOW_ZERO
0.00
O/S
39 2039
LOW_MF
1.0000
O/S
29 2029
1061:
m/s
O/S
27
2027 LIN_TYPE
1: linear with input
O/S
PRIMARY_VALUE_
FTIME
NOMINAL_SIZE_
UNIT
SECONDARY_
VALUE_UNIT
SECONDARY_
VALUE
SELECT_FLOW_
TUBE
PRIMARY_VALUE_
LOWCUT