IM 01E20F02-01E
5-3
5. EXPLANATION OF BASIC ITEMS
DISPLAY_SELECT1, 2, 3:
Table 5.3.1 DISPLAY SELECT
DISPLAY_SELEC T 1
DISPL AY_SELE CT 2
DISPL AY_SELE CT 3
The display content for the display
unit's first line.
The display content for the display unit's
second line.
The display content for the display
unit's third line.
1:Flow Rate(%)
1:Off
1:Off
2:Flow Rate
2:Flow Rate(%)
2:Flow Rate(%)
3:Integrator1 Out
3:Flow Rate
3:Flow Rate
4:Integrator2 Out
4:Flow Rate(Bar)
4:Flow Rate(Bar)
5:Arithmetic Out
5:Integrator1 Out
5:Integrator1 Out
6:Integrator2 Out
6:Integrator2 Out
The factory default setting is 2.
7:Arithmetic Out
7:Arithmetic Out
8:PD Tag
8:PD Tag
9:Adhesion Check
9:Adhesion Check
10:Communication
10:Communication
The factory default set ting is 1.
The factory default setting is
T0501.eps
1.
DISPLAY_CYCLE:
Sets the cycle of LCD display. The factory default
setting of the display cycle is 2: 400ms. The valid
range can be selected from below;
1: 200ms
2: 400ms
3: 1s
4: 2s
5: 4s
6: 8s
If the low temperature environment makes it difficult
to view the display, it is recommended that you set a
longer display cycle.
PRIMARY_VALUE_TYPE:
Indicates the measuring value type used in
PRIMARY_VALUE. Valid range are as follows;
100: mass flow
101: volumetric flow
102: average mass flow
103: average volumetric flow
65535: other
Factory default is 101: volumetric flow.
ALARM_PERFORM
(1)
Overview
This parameter masks Alarm/Warning. By
setting "0" to each bit, corresponding
Alarm/Warning are cleared. When masked the
corresponding bit of DEVICE_STATUS
becomes OFF and no alarm is displayed on
LCD, and also becomes out of scope of
Primary value status, ED_ERROR setting.
(2)
Bit mapping (0 :MASK, 1 :NON MASK)
bit
bit
categorize
categorize
Alarm
Alarm
0
Process
30:Sig Overflow
1
Alarms
31: Empty Pipe
2
Warning
33:Adhesion Alm
3
4
80:Adhesion Wng
82:Auto Zero Wng
defalt
defalt
1
1
0
0
1
5
85:Flow Vel Over
1
6 ~ 15 Not used in AXF
0
23 ~ 31 Not used in AXF
0
16
AI
42:AI FB O/S Mode
110:AI Lo Lo Alm
111:AI Hi Hi Alm
130:AI Non-Schedule
141:AI Sim. Enabled
150:AI FB Man Mode
1
17
IT1
43:IT1 FB O/S Mode
131:IT1 Non-Schedule
151:IT1 FB Man Mode
120:IT1 Low Clock Per
0
18
IT2
44:IT2 FB O/S Mode
132:IT2 Non-Schedule
152:IT2 FB Man Mode
121:IT2 Low Clock Per
0
19
DI1
45:DI1 FB O/S Mode
133:DI1 Non-Schedule
142:DI1 Sim. Enabled
153:DI1 FB Man Mode
0
20
DI2
46:DI2 FB O/S Mode
134:DI2 Non-Schedule
143:DI2 Sim. Enabled
154:DI2 FB Man Mode
0
21
AR
47:AR FB O/S Mode
135:AR Non-Schedule
155:AR FB Man Mode
122:AR Range Set Err
0
22
PID
42:PID FB O/S Mode
112:PID Lo Lo Alm
113:PID Hi Hi Alm
136:PID Non-Schedule
156:PID FB Man Mode
160:PID FB Bypass Mode
0
T0502.eps