![YOKOGAWA 735201 User Manual Download Page 53](http://html.mh-extra.com/html/yokogawa/735201/735201_user-manual_911162053.webp)
Chapter 5 Notes When Purchasing the Sensor Head by Itself
IM 735201-01E
5-1
5.1 Uploading Sensor Specific Data
In order to satisfy sensor head operating characteristics, sensor specific data is
uploaded. This data is included in the TB200 Utility CD that comes with the
sensor head. Upload the data using your PC and USB cable. Note that if you
selected the combined calibration option at the time of purchase (suffix code
-CA1 or -CA3), the sensor specific data (calibration data) has already been
loaded into the TB200 Optical Powermeter, so this step is not necessary.
In the following cases, sensor-specific data must be uploaded before starting
measurement.
●
When using a sensor head purchased by itself as an add-on.
●
If the suffix code printed on the name plate on the rear panel of
the TB200 is -CA0.
Preparation
1. Install the USB driver (included) on the PC (see section 4.2 for the
installation procedure).
2.
Turn the instrument ON, then connect the instrument to the PC using a
USB cable (insert the cable connector completely and firmly into the
receptacle). The instrument automatically enters USB mode, and USB is
displayed (see section 2.3).
Upload Procedure
1.
Place the TB200 Utility CD in the CD-ROM drive.
2. Choose
My Computer
>
CD-ROM
>
Utility
.
3. Double-click
the
TB200DataUpLoad.exe
file in the Utility folder.
The TB200DataUpLoad screen is displayed.