37
IM 702915-01EN
4. Product Specifications
Electrical Specifications
Accuracy value is for 23 ±3 °C, 30 minutes after power-on.
Item
702915
702916
Frequency band
DC to 50 MHz (–3 dB)
1
DC to 120 MHz (–3 dB)
1
Rise time
7.0 ns
2.9 ns
Delay time
2
30 A range 12 ns (typical)
3
5 A range
12 ns (typical)
3
0.5 A range 13 ns (typical)
3
Current range
(Output voltage rate)
30 A range (0.1 V/A)
5 A range (1 V/A)
0.5 A range (10 V/A)
Continuous
Maximum
input
4
30 A range 30 Arms
5
5 A range
5 Arms
5
0.5 A range 0.5 Arms
5
Amplitude
accuracy
6
30 A range ±3.0% rdg. ±1 mV,
±1.0% rdg. ±1 mV (10 Arms or less, typical)
3
5 A range
±3.0% rdg. ±1 mV,
±1.0% rdg. ±1 mV (typical)
3
0.5 A range ±3.0% rdg. ±10 mV,
±1.0% rdg. ±10 mV (typical)
3
Maximum
peak current
30 A range ±50 Apeak (maximum 2 sec input)
5 A range
±7.5 Apeak
0.5 A range ±0.75 Apeak (less than 10 MHz),
±0.3 Apeak (10 MHz or more)
Noise
7
75 µArms, 60 µArms (typical)
3
Input impedance
Depend on frequency
8
Temperature characteristics
of sensitivity
9
±2.0% rdg.
Effects of radiated radio-
frequency electro-magnetic
field
±10 mA or less at 3 V/m
Effects of conducted radio-
frequency electro-magnetic
field
±10 mA or less at 3 V
Effects of external magnetic
field
10
20 mA or less
5 mA or less
1 Refer to "Frequency Characteristics" for frequency characteristics diagram.
2 Delay time for input signal 1ns rising waveform.
3 "Typical" values are typical or average values and are not strictly guaranteed.
4 Specified by DC and sine wave. Depending on the ambient temperature or
measurement conditions, the internal temperature may increase and these
values may decrease.