26
IM 702915-01EN
4.
Turn on the waveform measuring instrument with a probe power
supply or the 701934 power supply. The instrument’s power lamp
(POWER) illuminates, and the DEMAG/AUTO ZERO LED blinks
slowly while demagnetization and zero adjustment is taking place.
Note
• The instrument’s output is internally terminated. Use a waveform
measuring instrument with an input impedance of 1 MΩ or higher.
Correct measurement is not possible with an input impedance of 50 Ω.
• The offset drift may be large immediately after starting to supply
power due to instrument’s self-heating and other effects. To make
accurate measurements, warm up for at least 30 minutes.
3. Operation Procedure