yellow jacket 40870 User Manual Download Page 5

 

 

 

 

 

Do not operate the instrument around 

explosive gas, vapor, or dust. 

 

 

Various refrigerants have been intentionally 

excluded for very significant safety reasons. 

Never use refrigerants in this instrument 

that are not listed in the Set-up menu. 

 

 

The refrigerant database in this unit may 

include refrigerants classified as flammable. 

If such refrigerants are selected, the 

operator may need additional certifications 

and/or training. Consult your government 

and legal authority and comply fully with all 

requirements. 

 

 

Always wear eye and skin protection when 

working with refrigerants. Escaping 

refrigerant vapors will present a freezing 

danger. Do not direct refrigerant vapors 

venting from hoses towards the skin. 

 

 

Maximum Working Pressure: 700 psia (4.83 

MPa) 

 

 

Because this instrument allows for various 

inputs including electrical and mechanical, 

care must be taken to observe any ways 

that an electrical shock hazard could 

develop. Example: Wet or humid conditions, 

along with a damaged thermocouple or 

vacuum sensor, could allow an electrical 

path across the instrument and over wet 

hoses. Keep all interconnected equipment 

clean, organized, and in proper condition. 

Do not use the instrument if you are not 

qualified to recognize potential electrical 

faults. 

 

 

 

 

 

 

To avoid damage to equipment, follow these 

guidelines: 

 

 

Do not allow pressures beyond the 

specifications listed in this manual. 

 

 

Be aware that internal pressures can change 

unintentionally when equipment is stored 

with pressure in the system during 

temperature changes. If sub-cooled liquid 

refrigerant is trapped in a hose or manifold 

with no room for expansion, it may result in 

dramatic pressure variations with seemingly 

small temperature changes. Pressures can 

reach high enough levels to cause damage 

to the instrument’s internal pressure 

transducers. Release liquid refrigerant from 

the hoses and manifold when disconnecting 

from a system. 

 

 

Do not attempt to introduce liquid or 

samples heavily laden with oil into the 

instrument. 

 

 

Do not use this instrument on systems 

containing leak sealing chemicals. These 

leak sealants can collect and harden in the 

instrument, causing permanent damage. 

 

 

 

 

 

 

Summary of Contents for 40870

Page 1: ...ese Instructions do not cover the manifold attached to the instrument For instructions on use of the TITAN 4 Valve Manifold please visit https yellowjacket com product titan 4 valve test and charging...

Page 2: ...des Menu 9 Pressure Temperature Mode 9 Overview 9 Interpreting the Pressure Analog Gauges 10 Interpreting the Pressure and Temperature Line Graphs 10 Operating Pressure Temperature Mode 11 Evacuation...

Page 3: ...e Units 22 Temperature Units 22 Vacuum Units 22 Device Settings Menu 22 Brightness 22 Backlight Timer 22 Auto Off 23 Logging Rate 23 Ch 5 YJACK View App Integration 24 Overview 24 Operation of the YJA...

Page 4: ...oid personal injury or death follow these guidelines Most governments and legal authorities re quire that HVAC technicians be trained and certified in the safe and proper operation of HVAC tools such...

Page 5: ...damaged thermocouple or vacuum sensor could allow an electrical path across the instrument and over wet hoses Keep all interconnected equipment clean organized and in proper condition Do not use the i...

Page 6: ...and release the power button located at bottom center of the display The YJ logo will appear briefly on startup The current firmware version and Serial Number are displayed at the bottom left and righ...

Page 7: ...tions on touchscreen calibration see page 21 Connecting and Using the Temperature Clamps Two 67010 temperature clamps pictured in Figure 6 are supplied with the P51 870 TITAN and can be used to monito...

Page 8: ...distance from the vacuum pump such that it does not disturb the vacuum measurements Interpreting the Battery Life Indicator This instrument utilizes a 2000mAhr rechargeable lithium ion battery It is e...

Page 9: ...s equipped with an adjustable backlight and backlight auto dimming capabilities When the backlight dims during use a touch anywhere on the screen or a quick press of the power button can be used to wa...

Page 10: ...nvention preventing duplicates See Figure 15 for example of stored datalog files Ex 50th week of 2017 with device serial number ending in 127 4th datalog file 750127AD These files cannot be modified w...

Page 11: ...to a larger time increment the LED will not flash blue as often Flashing Red When the device reaches low charge or falls below 10 battery charge the LED will flash red every time a datalog sample is...

Page 12: ...icon in the upper right hand corner Pressure Temperature Mode Overview The P51 870 TITAN Pressure Temperature mode can be used to accurately diagnose and service any system with compatible refrigerant...

Page 13: ...s displayed in a digital format at the center of each blue and red analog gauge respectively Each gauge needle adjusts in real time and can be used to observe fluctuations in pressure as if utilizing...

Page 14: ...e at any time by tapping anywhere within the graph Operating Pressure Temperature Mode While on any modes screen or the modes menu tap the settings icon in the upper right hand corner to access the ge...

Page 15: ...l Settings Button System vacuum pressure is displayed in real time as a digital representation as denoted by Pvac analog as represented by the green analog gauge and a graphical format as a logarithmi...

Page 16: ...ding Use the Target and Hold buttons to set the target vacuum level and vacuum hold timer respectively Pressing each of these buttons will cycle between a range of available values with the last value...

Page 17: ...igerant and non condensables can cause the vacuum level to rise in a system giving a false positive during a leak test when no leak is present Pressure Hold Mode Pressure Decay Test Overview The P51 8...

Page 18: ...ear scale with minor divisions displayed as small dots and major divisions displayed as larger dots respectively The current system pressure is displayed in a digital format at the center of the analo...

Page 19: ...s measured as a percent of the initial system pressure the system can drop before the P51 870 TITAN deems the pressure decay test a failure To adjust the allowable change percent drag the slider bar u...

Page 20: ...will display a FAIL notice Figure 24 and the buzzer will sound indicating the test is complete To silence the buzzer exit the pressure decay test by selecting the modes button If the pressure decay t...

Page 21: ...gital analog vacuum gauge as evacuation mode The pressure rise gauge operates on a logarithmic scale from atmosphere ATM to 0 microns A digital reading and currently selected vacuum units are displaye...

Page 22: ...desired settings have been set select the continue button to advance to the pressure rise test The back button can be used to return to the previous evacuation screen After selecting the continue but...

Page 23: ...gs menus tap on the appropriate button The exit button in the bottom left corner can be used to return to the previous screen at any time Zeroing the Pressure Transducers To ensure accurate pressure r...

Page 24: ...rigerant Settings Menu The refrigerant settings menu shown in Figure 31 can be accessed either by tapping on the Change Refrigerant button in the general settings menu or by tapping the button in the...

Page 25: ...the Device Settings button in the general settings menu To adjust the display brightness tap and drag the slider bar next to brightness until the desired brightness is displayed To adjust the backligh...

Page 26: ...t the logging rate to the lowest possible time interval 1s but this may result in extremely large datalog files making it difficult to analyze the information Be sure to set the logging rate to an app...

Page 27: ...ce Navigate to the YJACK VIEW App and make sure the manifold is powered on The P51 870 TITAN should appear in the available devices menu as the name of the device followed by the serial number Ex YJP5...

Page 28: ...dirty generously soak a soft cloth with warm soapy dish soap water and place the cloth for a couple of minutes over the display to loosen any stubborn dirt Wipe off excess water with a clean less damp...

Page 29: ...acklight Size Approx 7 5 x 8 5 x 4 25 Weight 2 63 lbs Table 7 2 Instrument Specifications Working Pressure 0 700 psia 48 3 bar Pressure Sensing Resolution 0 1 psi 0 1 bar 1 kPa 0 001 MPa 0 01 kg cm 2...

Page 30: ...d Temperature clamp cable damaged Call technical support Temperature clamp jacks damaged Call technical support Clamp not properly attached to system Check clamp connections to system Low Side and Hig...

Page 31: ...28...

Reviews: