table 4 electromagnetic immunity
|
Table 4: Immunity (Stimulation mode) table according to IEC 60601-1-2, 3
rd
edition
Guidance and manufacturer’s declaration – electromagnetic immunity
The ARYS™ me is intended for use in the electromagnetic environment specified below. The customer or user of the
ARYS™ me should ensure that it is used in such an environment.
Immunity test
standard
IEC 60601
test level
Compliance level
Recommended separation distance
c
Portable and mobile RF communications
equipment should be used no closer to
any part of the ARYS™ me, including
cables, than the recommended separa-
tion distance calculated from the equa-
tion applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Not applicable
d
= 1.2 √
P
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
10 V/m
80 MHz to 2.7 GHz
d
= 0.35 √
P
80 MHz to 800 MHz
d
= 0.7 √
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufactur-
er and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromagnet-
ic site survey,
a
should be less than the
compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol: