background image

                              

 

 

 

 

ONE-TOUCH WIRELESS HEALTH TRACKER 

 

BODY LIFE TRACKER

 

AMI Spirometer 

User Manual 

Model: VA08 

 
 
 

                                                         

 

 

 

 

Document No  : AMI-PRD-UM-B-VA08 

Version  

:  1.0 

Date 

 

:  29.12.2015 

Summary of Contents for VA08

Page 1: ...ONE TOUCH WIRELESS HEALTH TRACKER BODY LIFE TRACKER AMI Spirometer User Manual Model VA08 Document No AMI PRD UM B VA08 Version 1 0 Date 29 12 2015 ...

Page 2: ...trends India Private Ltd be held liable for any loss expenses or damage of any kind whatsoever whether direct indirect incidental or consequential arising from the design or use of this product or the support materials provided with the product Limited Warranty No warranties are made either expressed or implied with regard to the contents of this work its merchantability or fitness for a particula...

Page 3: ...oduction 4 3 0 Package Contents 5 4 0 Instructions for safe use 5 5 0 Intended use 6 6 0 Prerequisites 6 7 0 Using Mobile Application 6 8 0 AMI Spirometer Measurement procedure 6 9 0 Charging the AMI Spirometer 7 10 0 Explanation of Indicators Symbols 9 11 0 Technical Specifications 10 12 0 Certifications 11 13 0 FCC compliance statements 12 Contact us 13 ...

Page 4: ...e lung function This device is used to assess the condition of the lungs by measuring the volume of air inhaled and exhaled The system consists of the electronics part with enclosure and a disposable turbine where the user blows and sucks air The device calculates the flow rate and other relevant parameters and sends the data wirelessly to the mobile device The spirometer device is powered by 5V D...

Page 5: ...test for another user Do not use unauthorized parts and accessories Do not store or use this device in locations which are maintained above the specified temperature and humidity ranges Do not use in an operating theatre nor in the presence of inflammable liquids or detergents nor in the presence of inflammable anaesthetic gases oxygen or nitrogen Do not expose the turbine to a direct jet of water...

Page 6: ...computer mobile device should be 7 or higher The Windows OS version of the using computer mobile device should be 8 or higher 7 0 Using Mobile Application Downloading and Installing Mobile application in your computer mobile device 1 Download and Install the application from Play store Application store Windows application market 2 On successful installation launch the application in the mobile de...

Page 7: ...the teeth being carefully to ensure that air cannot escape from the sides of the mouth 11 Exhale completely and followed by complete inhale 12 If the measurement was successful the application would display the reading successfully If the reading was interrupted or if the user has not used full effort then the app would ask the user to re take the reading 13 Remove the mouth piece from the mouth a...

Page 8: ...SB cable AMI Spirometer Charging Procedure 1 Connect the USB cable to the power adapter 2 Connect the micro USB port to the spirometer power port USB port 3 Plug the adapter to the AC mains 110V 60Hz or 230V 50Hz 4 Switch ON the AC mains to charge the device 5 Check whether the power indicator turns red while charging the device ...

Page 9: ...ng with the mobile tab o Steady Bluetooth light indicates that the gadget is paired and ready for the measurement o Steady green light indicates that the gadget is operating in battery mode o Blinking of green light indicates that the gadget has low battery level Steady red light indicates that the gadget is in the charging mode Symbol for ENVIRONMENT PROTECTION Symbol for KEEP DRY Symbol for MANU...

Page 10: ...pirometer Unit Dimensions 145 H x 45 W x 10 T mm Unit Weight 100 grams Including Battery Power Source DC 5V 2A Bluetooth Bluetooth 4 0 EDR or Lower Bluetooth Low Energy BLE Operating Temperature 16 C to 35 C 60 8 F to 95 0 F Storage and Transport Temperature 10 C to 55 C 14 F to 131 F Storage Humidity 30 RH to 90 RH Volume Accuracy 3 or 50mL Flow Accuracy 5 or 200 mL s Flow Range 16 L s NOTE These...

Page 11: ...dical devices Part 10 Tests for irritation and skin sensitization EN ISO 80601 2 61 2011 Particular requirements for basic safety and essential performance of pulse oximeter equipment EN 80601 2 30 2010 Particular requirements for basic safety and essential performance of automated non invasive sphygmomanometers EN 12470 5 2003 Clinical thermometers Part 5 Performance of infra red ear thermometers...

Page 12: ... limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy And if not installed and used in accordance with the instructions may Cause harmful interference to radio communications However there is no Guarantee that interference will not occur in a particular installation If th...

Page 13: ...MARAN NAGAR SEMMANCHERY VILLAGE CHENNAI TN 600119 Tel 91 44 66540922 Fax 91 44 66540901 Email support amibolt com Website www amibolt com www amiindia co in American Megatrends International GmbH Truderingerstraße Truderinger Strasse 283 81825 München Munich Deutschland Germany Tel 49 89 969995 10 Support 49 89 969995 11 Fax 49 89 969995 19 Email europesales ami com ...