| table 2 electromagnetic immunity |
Table 2: Immunity (Stimulation mode) table according to IEC 60601-1-2, 3
rd
edition
Guidance and manufacturer’s declaration – electromagnetic immunity
The ARYS™ me is intended for use in the electromagnetic environment specified below. The customer or user of
ARYS™ me should ensure that it is used in such an environment.
Immunity test
standard
IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
Excessive electrostatic discharge may lead
to an error on the tracker. Should this
happen, please proceed according to the
troubleshooting table on page 53.
Electrostatic fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Not applicable
–
Surge
IEC 61000-4-5
± 1 kV line to line
± 15 kV line to earth
Not applicable
–
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % U
T
(0.5 cycle)
40 % U
T
(5 cycles)
70 % U
T
(25 cycles)
<5 % U
T
for 5 s
Not applicable
Note: U
T
is the AC mains voltage prior to application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A /m
Not applicable
Not applicable according to IEC 60601-1-2
4
th
ed. for non-magnetic field sensitive
devices