10
Operation Method
Sequence of Operation
CAUTION!
It is possible to make vibrate the main unit, when the setting sample is
heavy or the shaking time is large. Please refer to the table of
vibration-proof and operate not to move by sample or shaking time.
The case of horizontal shaking type
The sample weight that can be shaken in each option is as follows.
Erlenmeyer Flask Holder
(
Item Code:232097
)
Shaking time (Scale)
1 2 3 4 5 6 7 8 9 10
Permissible sample
weight (kg)
5.0
5.0
5.0
5.0
5.0
5.0
5.0 5.0 2.5
1.5
Test Tube Holder(Item Code:232086)
Shaking time (Scale)
1 2 3 4 5 6 7 8 9 10
Permissible sample
weight (kg)
1.0
1.0
1.0
1.0
1.0
1.0
1.0 1.0 1.0
×
Test Tube Rack Holder(Item Code:232088)
Shaking time (Scale)
1 2 3 4 5 6 7 8 9 10
Permissible sample
weight (kg)
5.6
5.6
5.6
5.6
5.6
5.6
5.6 3.5 1.0
×
Centrifugal Tube Holder(Item Code:232087)
Shaking time (Scale)
1
2
3
4
5
6
7
8
9
10
Permissible sample
weight (kg)
2.5
2.5
2.5
2.5
2.5
2.5
2.5 2.5 × ×
Two Layers Spring Shaking Rack (Item Code:232079)
Shaking time (Scale)
1
2
3
4
5
6
7
8
9
10
Permissible sample
weight (kg)
15
15
15
15
15
15
15 × × ×
×…No Shaking
CAUTION!
・
The holder or the flask are not attached.
・
It might be possible to shift the permissible sample weight by the
condition of setting floor.
・
Permissible sample weight is the value with maximum load weight.
CAUTION!
In the early stage after supplying power, the shaking times may be
10-15% lower than the set value. Perform the running operation for
about 15 min. Then, the shaking times will stabilize.