Top Screen
Connection Setting Screen
TAKING MEASUREMENTS
Follow the procedure outlined below for taking basic measurement readings.
1. Turn on unit power. The status light will turn red
●
.
* If battery level falls too low, the red light will begin blinking.
2. Open the measurement chamber lid and set PCR tube into the measurement chamber.
Firmly close the lid.
3. Launch the tablet application
(PAS-110) by tapping the
icon.
Page10, 1. Using the Top Screen.
4. Make BLE connection.
Page11, 2. Wireless Connection.
Note that the tube lid
connector should fit snugly
between the two guide
projections at the back of
the chamber.
CAUTION
Do not let any sample liquid from the PCR Tube spill into the unit.
This could damage the optical unit.
Guides
Connector
TAK
IN
G
ME
ASU
REME
NTS
8