Illustration 3-7:
Selection of Concentration Units
Illustration 3-8:
Results of Standard Curve
Measurement Procedure
TAKING MEASUREMENTS
(2)
Tap the
①
Concentration Units Select Button
to
choose the units in which the measurements will be
made (Illustration 3-7). Tap the OK to confirm.
Initially, four unit variations are registered (mg/mL, μg/
mL, ng/mL, and %). See
<Pages 20-22>
for how to add
new units.
(3)
Tap the
②
Known Concentration Input Area
and
input the known concentration value.
(4)
Tap th e
③
K n ow n Co n ce ntrati o n S am p l e
Measurement Intensity Display
to display the black
frame.
(5)
Tap the
④
Measure Button
to begin measurement.
The measured value will be displayed in white font.
(6)
The
⑤
Absorbance Value Display
will show the
actual calculated absorbance.
(7)
When this measurement is finished, remove the cur-
rent sample in the photo absorbance sensor, place
the next sample in, return to step ( 1 ), and repeat the
process.
(8)
When all the samples have been measured, tap
the
⑥
Standard Curve Graph Display Button
and
confirm the graphed results. The values displayed
in white font (Illustration 3-8) are the ones displayed
on the standard curve
<Page 18>
. If you wish to a
measurement from the graph, remove it by a long
press on that data’s display box.
(9)
When you return to the top screen and proceed to
“Measure”, the unknown sample is measured using
the created standard curve
<Page 25>
. The standard
curve data is stored together with the measurement
data
<Page 34>
.
* The standard curve cannot be saved by itself.
TAK
IN
G
ME
ASU
REME
NTS
17