66
10. Specifications
Model
INE800
System
Forced blow circulation
Operating temperature range *1
0
℃~
60
℃
Temp. control accuracy
(
at 37
℃)
*1
±
0.2
℃
former JTM(continuous run at
37
℃
)
±
0.3
℃
JIS Temp. fluctuation
(continuous run at 37
℃)
±
0.5
℃
former JTM(cycle run at 37
℃
)
±
1.0
℃
JIS Temp. fluctuation (cycle run
at 37
℃
)
Temp. distribution accuracy
(
at
37
℃)
*1
±
0.5
℃
former JTM(continuous run at
37
℃
)
2.0
℃
JIS Temp. gradient (continuous
run at 37
℃
)
Temp.
attainment
time
*1
Max. temp.
20
℃→
60
℃
Approx. 35 minutes
Min. temp.
20
℃→
0
℃
Approx. 50 minutes
Temp. control system
PID control
Temp. setting system
Set temp. display
:
5-digit amber LED digital display
Temp. display system
Temp. display
:
4-digit green LED digital display
Timer
0min
~
99 hr 59min, un-limited
Resolution of the timer
One minute
Operating function
Fixed temperature operation: auto stop; auto start
Program run
(
Max. 99 patterns, 99 steps, repeat
)
Additional functions
Integral time meter function, calibration offset function
External communication function (RS485), temp. output terminal
(
4-20mA
)
Refrigerator
Rotary unit 200W
Operating range of refrigerator
Setting 40
℃
or lower
※
2
Refrigerator refrigerant
R134a
350g
Heater
Material
Iron chrome wire
Capacity
750
W
Blower fan
Axial fan
Sensor
Pt100
Ω
(temp. controller), K-thermocouple(standalone overheat preventive device)
Interior
Stainless steel plate
Exterior
Chrome free electrogalvanized steel plate Chemical-proof baking finish
Heat insulator
Styrene foam (non-Freon)
Inner door
Reinforced glass (thickness:5
㎜
; two-part of upper and lower
)
Safety unit
Temp. sensor error, heater disconnection, SSR short-circuit, main relay error, internal
fan error, refrigerator high voltage error, refrigerator discharge gas temperature error,
inverter error, automatic overheat preventive function, key lock function, standalone
overheat preventive device, overcurrent ELB
Defrosting mechanism
Manual ON/auto OFF, auto defrost
(
time-dependent cycle or clock time
)
Cable port
I.D.:50
㎜(
right side of the main unit
)
Internal dimensions:
W × D × H
*3
600
×
477
×
1000
(
effective size:800
)
mm
External dimensions:
W×D×H
710
×
645
×
1730 mm
Inner capacity
286L
Withstand load of shelf board
15
㎏
/board
Number of shelf stages (pitch
)
23
(
30
㎜)
Power supply
AC100V
10A
(
ELB: 15A
)
Weight
Approx.135
㎏
Accessories Shelf
boards/shelf
rails
Stainless steel punched metal 5 boards/10 rails
Others
Door key x 2, silicon plug x 1
Optional accessories
External communication adaptor set(RS-485-USB conversion), external alarm
terminal, shelf board set, time-up output terminal, anti-seismic clamps (anchor bolt
hole:
Φ
11mm)
*1 Performance has been measured at: power of AC100V; room temperature of 23
℃±
5
℃
; humidity of
65
%
RH
±
20
%
; and no load.
The recommended operating environmental temperature range:5
℃~
35
℃
.
*2 The refrigerator will turn OFF at a set temperature of 40.1
℃
or over. If you set cooling at a constant
rate in the program run, the refrigerator will be OFF and cooling at a constant rate is disabled when
the indicated temperature is 45
℃
or over and the set temperature is 40.1
℃
or over.
*3 Protrusions are excluded.