Yamatake Corporation
5-66
ST3000 Smart Transmitter Series 900 Electronic Differential Pressure/Pressure Transmitter
5-13-1 :Changing Mode of Operation
Procedure
If you need to change your transmitter’s mode of operation, use the steps in the follow-
ing table to change the mode from analog to digital or digital to analog.
Step
Description
SFC screen
1
Make sure that the transmitter is
in the ready state.
2
Press the
key and then press
the
key.
When the message to the right is
displayed, press the
key.
3
Press the
key for verifica-
tion purposes.
4
The procedure is completed when
these messages are displayed.
DSTJ PT001
READY...
SHIFT
Q
A
DE
ENTER
( Yes )
NON-VOL
SHIFT-
SHIFT-
DSTJ PT001
CHANGE TO DIGITAL?
ENTER
( Yes )
NON-VOL
DSTJ PT001
ARE YOU SURE?
DSTJ PT001
WORKING...
DSTJDE PT001
DEXMTR
DSTJ PT001
READY...