Commissioning
NO 800
Commissioning
Combination elec-
trode NO 800
Conditioning, calibration, measurement
General informa-
tion
When operating the electrode ensure that
the filling opening (1) for the bridge electrolyte is open
the inner junctions (3) are covered with bridge electrolyte
no air bubbles are in the bridge electrolyte
the depth of immersion is within the optimum range:
Before measuring
1
Remove the storage vessel.
2
Fill ELY/BR/503/N bridge electrolyte into the filling opening so that
the inner junctions are covered with bridge electrolyte.
3
Rinse the combination electrode with deionized water.
4
Wipe the shaft using a clean paper towel.
Minimum depth of
immersion
The outer junction (4) must be covered
Maximum depth of
immersion
Approx. 1 cm below the fluid level (2) of the
bridge electrolyte
1
2
3
4
1
Condition every ion sensitive electrode in standard solution before
using it.
Summary of Contents for wtw NO 800 DIN
Page 2: ...2 ba15362d04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...ba15362d04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15362d04 12 2021 Kalibrierkurve einer Nitrat selektiven Elektrode NO 800...
Page 15: ...OPERATING MANUAL ba15362d04 12 2021 NO 800 DIN NITRATE SENSITIVE ELECTRODE OPERATING MANUAL...
Page 16: ...ba15362d04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 23: ...NO 800 What to do if ba15362d04 12 2021 23 Cable length Plug DIN plug 1 m...
Page 24: ...24 ba15362d04 12 2021 What to do if NO 800...
Page 25: ...MODE D EMPLOI ba15362f04 12 2021 NO 800 DIN LECTRODE S LECTIVE DE NITRATE MODE D EMPLOI...
Page 26: ...ba15362f04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 34: ...34 ba15362f04 12 2021 Courbe de calibration d une lectrode s lective de nitrate NO 800...
Page 36: ...ba15362s04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 44: ...44 ba15362s04 12 2021 Especificaciones t cnicas NO 800...
Page 45: ......