Drinking water panel
Drinking water panel
Mounting
Ambient
conditions
Permissible over
-
pressure
Medium overflow Into open channel flow, virtually pressureless. A water column of max. 0.5 m
relating to the lid of the pH flow through armature may be applied.
Sample flow
Range of application Drinking water and swimming pool water
Inlet and outlet
Dimensions (H x W x D)
Approx. 850 x 700 x 13 mm
Material
PVC rigid foam, spumed, white
4 holes in the drinking water panel
Diameter
8.5 mm
Distances
60 mm from the edge
Temperature
Operation
5 °C ... + 40 °C
Storage
0 °C ... + 45 °C
Before pressure regula
-
tor (inlet)
0 ... 6 bar
After pressure regulator
1.5 bar (preset)
Without chlorine sensor
At least 10 l/h
With chlorine sensor
At least 15 l/h
Inlet
Tube nozzle R1/2", not removable
Outlet
Outlet into funnel, Ø 40 mm