134
4.6.8
Documentation
Three different format settings are available for documentation (Fig. 115) on a printer or USB device: «
short
»,
«
standard (with curve)
» and «
GLP
Fig. 115
Fig. 116
Method type
Short documentation
Standard documentation
GLP-documentation
Automatic
titration
Method name, date, time, duration of
titration, sample description,
weight/volume, starting and end
measurement values (pH/ mV Temp),
slope and zero point of the pH electrode,
results and calculation formula
Same as ‘Short
documentation’ +
titration curve
Same as ‘Standard
documentation’ +
method contents
Manual titration
Method name, date, time, sample
description, sample weight/sample
volume, results and calculation formula
N/A
Same as ‘Short
documentation’ + plus
method contents
Dosing
Method name, date, time
N/A
N/A
.