Fig. 62
Working point
The working point in mV is the base value for the indicator electrode. It can be set from 1 to 1000 mV. 300 mV is
pre-set as standard value and works for many reagent/reagent combinations. A higher value is maybe sometimes
necessary.
Fig. 65
Stop Drift (delta)
The stop drift in µg/min is not an absolute end value. The end value of a titration is always the currently measured
drift + stop drift.
Fig. 64
An example:
If the current drift shows a value of 1.5 µg/min and the stop drift is set to 2.0 µg/min, then the end drift actually to
be reached is 3,5 µg/min.