88
Fig. 51
The calculation of the mean value is already possible from two individual values. The calculation of the relative
standard deviation is only possible from 3 single values. The maximum quantity is 10.
Fig. 52
The mean value and the relative standard deviation (RSD) are shown directly on the display or in the result
printout.
4.6.1.7 Global Memories
Results of titrations can be written into one of the 50 global memories (M01 - M50) for additional calculations.
Fig. 53
The mean value is written into the global memory when the statistic is switched on. You enter the submenu with
<Enter/OK>
. If a global memory has not been created, a memory can be created by using the insert key
<Ins>
.
The titrator proposes a memory name, such as
M01
(M01- M50). The name of the memory can be changed in
reference to the application. Here in this example of
“M01”
for
“blank value extern…”
.
Summary of Contents for SI Analytics TITRATOR TitroLine 7500 KF
Page 1: ...TitroLine 7500 KF trace TITRATOR OPERATING INSTRUCTIONS...
Page 57: ...Notes...