Put all white inner plastic adapters to the waste, solvent and moisture bottle. Fill the moisture bottle with
molecular sieve. Connect the PVC and PTFE plastic tubes as shown in the next pictures.
The PVC tubes are connected to the connectors at the back side of the TM 235 KF. The long PVC tube is used
for the connection of the waste bottle. The two shorter PVC ones are used to connect the moisture bottle and the
solvent bottle. The moisture bottle is connected to the right connector (view from above) of the TM 235 KF. The
waste (clear) bottle is connected to the left connector.
Fig. 8
Put the threaded pipe with the NS 14/23 core and the GL 14-thread in one of the NS-14.5-openings. Put both
PTFE-tubes through both bore holes of the septum. The PTFE tube from the clear waste bottle is adjusted to the
bottom of the titration vessel (tube 1). Put the PTFE tube from the solvent bottle (tube 2) is adjusted as shown in
the next two pictures:
Fig. 9