55
Dissolved Oxygen Transmitter
GB Oxix Manual 1209
FW: 841013-009, 846001-002, 841513-009
A typical cleaning interval for aeration basins would be 2 hours (120 minutes).
In colder climates, condensation may form and then freeze in the tubing. This
can be prevented by setting the interval to 5 - 20 minutes.
The cleaning time would typically be 30 seconds.
The measurement delay (recovery time) may also be used to coordinate with
other processes. The delay merely freezes the last reading before cleaning
begins. A typical value would be 60 seconds.
Averaging
Units
mA output
Converter setup
20 mA output
20 mA =
25.00 mg/l
Back Select
OK
D.O. Units
%
mg/l
ppm
Back Select
OK
Averaging
Units
mA output
Cleaning
High alarm
Low alarm
Sensor error
Status
Factory settings
Service menu
Back Select
OK
4mA output
4 mA =
00.00 mg/l
Back Select
OK
When ”n/a” is displayed, the
menu is not available or not
applicable
Yes
Enable mA output?
Yes
No
Back Select
OK
Averaging
0001 sec
(from 1 sec to 120 min)
Back Select
>
Temperature unit
°C
°F
Back Select
OK
Temperature
This menu is for display only.
Clean now
Yes
No
Back Select
OK
Reset Maintenance
Yes
No
Back Select
OK
Maintenance
Runtime 1000 hour
Count
500
Back Select
OK
Clean now
Select cleaning interval
060 Min
(20 - 255 Min)
Back Select
>
Cleaning
Interval
Use DI to clean
Clean now
Maintenance
Back Select
OK
Interval
Off
On
Back Select
OK
Use DI to clean
Interval
Maintenance
DI function
NO
NC
Back Select
OK
Use DI to clean
No
Yes
Back Select
OK
Delay
00100 ms
Back Select
>
Electro-mechanical relay
Opto relay
Not in use
Back Select
OK
Cleaning time
005 sec
Back Select
OK
Cleaning time
005 sec
Back Select
OK
Measurement delay
10 sec
Back Select
OK
GB Oxix menuer #1 Converter setup 1
FW: 841013-009, 846001-002, 841513-009
Page 2 of 7 26. september 2008