37
Dissolved Oxygen Transmitter
GB Oxix Manual 1209
FW: 841013-009, 846001-002, 841513-009
8. Startup
Initial Checks before Power-up
Before switching on power the following steps must be checked.
1. The local mains power supply voltage corresponds to the voltage printed
on the identification label of the transmitter.
2. All electrical connections are made in accordance with the electrical con-
nection diagram shown with special emphasis on the sensor connection
described on page
30
.
3. All terminal screws are tightened.
4. All cable glands are tightly secured.
Initial Power-up
1. Turn on power.
2. Within 60 seconds the Oxix
®
sensor will be found by the automatic sen-
sor detection program. During this process you might briefly see "SuSix1"
and/or "MagFlux1" on the display. This is normal and to be expected.
Initial Measurement
1. Make sure that the sensor is completely covered by liquid.
2. Turn on power to the transmitter and wait 15 minutes before taking the
initial measurement.
Language Selection
1. The default display language is English. If another language is required,
proceed with step 2.
2. Select "Display Setup" from the Main Menu.
3. Select "Language" from the Display Setup menu and chose the required
language (see page
73
).