background image

9   

UM Temp74 PT100 Vio EN rev.1 19.05.2023 

 

Replacing batteries  

The instrument runs on 3 AA 1.5V batteries. For replacement: 

1.

 

switch off the device; 

2.

 

turn the instrument downwards, with the connector part resting on a table top, so that the battery 

compartment cap is facing upwards, towards the user; 

3.

 

while holding the battery compartment cap with two fingers, completely unscrew the screw next to the battery 

symbol with a screwdriver; 

4.

 

remove the battery compartment cap; 

5.

 

remove the 3 spent batteries and insert new ones. Pay attention to correct polarity. Follow the diagram above 

the battery symbol in the rear compartment of the instrument; 

6.

 

reinsert the battery compartment cap; hold it in place with two fingers, insert the screw and screw in. 

 

 

Instrument transport  

The instrument always comes with its own carrying case. Only use the original carrying case to transport the 

instrument. Contact your local distributor if you need to buy it back. The inside of the case is shaped to 

accommodate the instrument and sensors. 

 

 

Input / Output Connections  

Only use original accessories guaranteed by the manufacturer. If necessary, contact your local distributor. 

 

 

 

 

 

 

 
 

 
 

 
 
 

Read the manual before connecting the probes 

 

Symbols and icons on the display  

                   

 

Symbol 

Description 

 

Error in measurement 

 

Battery charge indication 

 

Measurement Stability Indicator 

 

Automatic Datalogger 

 

Programmed Datalogger 

 

Number of data stored in data logger mode on instrument memory 

 

Press the arrow buttons to change the parameter or value on the display 

 

Instrument connected to the D software 

 

Reference channel 

 

Minimum and maximum values 

 

Measuring channel displayed 

 

Micro USB  

PC and Power supply connection 

CH4  

3 pin connector 

 for PT100 sensor 

CH2  

3 pin connector 

 for PT100 sensor 

 

CH3  

3 pin connector 
for PT100 sensor 

 

CH1 

3 pin connector 
 for PT100 sensor 

 

Summary of Contents for Temp74 PT100 Vio

Page 1: ...INSTRUCTION MANUAL Temp74 PT100...

Page 2: ...tilingual copy Espa ol Para reducir los gastos este manual ha sido impreso exclusivamente en italiano e ingl s Contacta tu distribuidor para recibir la copia digital multi idioma Deutsch Um umweltfreu...

Page 3: ...nstrumental Characteristics 6 Technical Data 6 Description Instrument 6 Display single channel 6 Multiview screen 7 Keyboard 7 Key Functions 8 LED 8 Installation 8 Supplied Components 8 Commissioning...

Page 4: ...ngs 14 Date and time 14 Units of Measurement 14 Language 14 Auto off 14 Sound 14 Password 15 Temperature measurement 15 Measure 15 MAX MIN function 15 HOLD function 15 Other functions 15 Automatic swi...

Page 5: ...ormation NOTE for useful product information Warning symbols Additional documents providing security information The following documents can provide the operator with additional information to work sa...

Page 6: ...mended to occasionally clean the instrument casing with a damp cloth and mild detergent This operation should be carried out with the instrument switched off and only by experienced and authorised per...

Page 7: ...MAX MIN reading Yes Offset Adjustment Yes for each channel independently Internal memory 200 000 data Data logger Manual Automatic Programmed Sampling 1 min 99 hours Working conditions Temperature 0 6...

Page 8: ...X and MIN values recorded Press for 3 seconds to activate and deactivate the HOLD function and to deactivate the MAX MIN recording function if active Directional key DOWN RM to scroll through menus an...

Page 9: ...he display brightness Only at this point do the buttons regain their function LED All instruments are equipped with a two colour LED red and green that provides the user with important information on...

Page 10: ...carrying case Only use the original carrying case to transport the instrument Contact your local distributor if you need to buy it back The inside of the case is shaped to accommodate the instrument...

Page 11: ...e below shows the Setup menu structure for the parameter for each programme the options the user can choose and the default value Description Options Factory Settings Sensor ID Probe 1 Probe 2 Probe 3...

Page 12: ...annels to be aligned to the reference channel Select our reference channel from the Setup menu Temp Settings Reference Probe and press to confirm In the Offset Probes menu select Align All The measure...

Page 13: ...the table above Note When the reset is complete the instrument returns to measurement mode and the parameters are reset to default settings Log Settings Menu TEMP Measurement Settings LOG Data logger...

Page 14: ...pendently once the set date and time is reached and will stop Note If recording takes place while there is a single channel measurement on the display data concerning only that channel will be saved I...

Page 15: ...er battery life Brightness Access this Setup menu to choose between three different display brightness levels Low Normal High Note Keeping the display at high brightness at all times negatively affect...

Page 16: ...appears on the display Always wait for the measurement to become stable Note If no temperature probe is connected the message CONNECT A PROBE appears When connecting or disconnecting probes the instru...

Page 17: ...corrected and the adjustment value with flashing cursor using the keys correct the value to Current Temperature to 0 0 C press to confirm and record the adjustment Note The adjustment affects the ent...

Page 18: ...e manufacturer will repair or replace defective components This guarantee applies only to the electronic part and does not apply if the product has been damaged misused exposed to radiation or corrosi...

Page 19: ...18 UM Temp74 PT100 Vio EN rev 1 19 05 2023 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax...

Reviews: